The VES-Matic 5 system: performance of a novel instrument for measuring erythrocyte sedimentation rate

  • Elisa Piva , Alice Stoppa , Michela Pelloso and Mario Plebani ORCID logo EMAIL logo

Abstract

Objectives

The VES-Matic 5 is an automated analyzer that assesses erythrocyte sedimentation rate based on a modified Westergren sedimentation technique. Instrument performance was established by addressing the recommendations of the International Council for Standardization in Haematology.

Methods

Comparison against the reference Westergren method was performed for all samples, and further for the low, middle, and upper third of the analytical range. Intra-run precision, inter‐run precision, and interference studies were further assessed. This study included the evaluation of reference ranges.

Results

The comparison of methods by Passing–Bablok analysis has shown a good agreement without systematic or proportional differences. The regression equation was y=−0.646 + 0.979x. The mean bias of −0.542 was obtained by Bland–Altman analysis and the upper limit of 8.03 with the lower limit of −9.11 can be considered clinically acceptable. Intra-run and inter-run precision were good for each parameter and interference studies did not show any significant bias with exception of anemia samples, which showed a proportional difference when comparing high erythrocyte sedimentation rate values. Using the local adult reference population, we verified the reference ranges in comparison to those available in the literature, and according to the Clinical Laboratory Standards Institute (CLSI) EP28-A3C document. We determined the upper limit partitioned by gender and the following age groups: from 18 to 50, from 50 to 70, and over 70.

Conclusions

The VES-Matic 5 analyzer presented good comparability with the reference method. As there are commercial quality control and suitable external quality assessment (EQA) material and programs, the VES-Matic 5 can be employed appropriately for routine purposes.

Introduction

The erythrocyte sedimentation rate (ESR) was applied in clinical practice following the studies by Edmund Biernacki in the 19th century, even though the test was standardized many years later by another scientist, Alf Vilhelm Albertsson Westergren [1]. Since its discovery, ESR is the laboratory test that quantifies a complex biological phenomenon, i.e., the behavior of a suspension of red blood cells in plasma under specified conditions. ESR is not linear and plotted against time; the sedimentation shows a sigmoid curve. Three phases are known. The first is the “lag phase” in which erythrocytes form rouleaux and aggregates and the sedimentation is very slow. The second, or “decantation phase,” is the true sedimentation phase. Erythrocytes fall more rapidly because the plasma interface and the sedimentation can occur more rapidly. In the final phase, erythrocyte aggregates pile up on the bottom of the tube [2], [3], [4]. New technologies and technical advances allow automated instruments to perform ESR testing, in consolidation with the modern organization of clinical laboratories. Automated instruments have great advantages like safety for operators, reduced turn-around time, and better analytic precision. Some of these instruments use the sedimentation principle by selecting an appropriate time interval, while others use other innovative techniques, which are different from the sedimentation rate [5].

For a long time, the International Council for Standardization in Haematology (ICSH) advocated ESR standardization and harmonization. The last document, published in 2017, was an obvious need to evaluate the different automated methodologies and all the working methods against that of the reference, that is the Westergren method. The recommendations were established to document the reliability, accuracy, and robustness of the results against the Westergren method for all new instruments, both for those based on the modified sedimentation principle and those that use different techniques [6].

Recently, a novel fully automated system, the VES-Matic 5 (DIESSE Diagnostica Senese, Monteriggioni Siena, Italy), has been proposed for ESR determination. Using closed ethylenediamine tetraacetic acid (EDTA) tubes and with a complete automation process, the instrument evaluates the sedimentation of the red cells and using a mathematical algorithm converts the raw data obtained into ESR results. The instrument also uses artificial intelligence (AI) to detect pre-analytical errors, for example, coagulated or unfilled samples. The aim of the present study was to perform the analytical validation of this new automated analyzer for ESR testing, in agreement with ICSH recommendations.

Materials and methods

Study design

The automated analyzer for ESR determination, Ves-Matic 5 was evaluated at the Department of Laboratory Medicine, University-Hospital of Padua according to the ICSH recommendations and included method comparison, precision study (intra-run and inter-run), hemolysis, triglycerides (TGs), and anemia interference studies, sensitivity to fibrinogen, and verification of reference ranges [6].

Specimens collection

Blood samples, used for method evaluation, precision determination, sensitivity to fibrinogen, and assessment of the interferences as well as verification of reference ranges, were selected from the daily routine. Samples were leftovers that had an initial request for cell blood counter (CBC) and/or ESR of both hospitalized patients and outpatients admitted to the Padua University Hospital. All samples were collected in 3.0 mL bi-potassium EDTA (K2-EDTA) BD Vacutainer® blood collection tubes (Becton Dickinson, BD-Plymouth, UK), processed according to manufacturers’ specifications, and analyzed within 5 h from venipuncture. The study followed the ethical principles for medical research involving human subjects, according to the World Medical Association (WMA) Declaration of Helsinki, and under the terms of all relevant local legislation.

VES-Matic 5

The new automatic instrument for the determination of the ESR, the VES-Matic 5 is based on a modified Westergren sedimentation technique. Briefly, all samples are collected in an EDTA anticoagulated tube, loaded into the instrument on the same rack of the CBC instruments, samples are mixed, and under controlled temperature evaluated for sedimentation measurement using an optoelectronic light source. The sedimentation rate is evaluated by a large number of optical recordings during a 20 min period and the differences are evaluated. Raw data are corrected for temperature variations and the obtained results are converted to Westergren values using a mathematical algorithm. The first result is performed after 28 min. Technologically innovative, the new analyzer applies an AI system in recognition of lipemic, hemolyzed, coagulated, or mislabeled samples, while the Internet of things, through an internal camera, is used for advanced remote assistance. In this way, the analyzer is connected directly to diagnostic devices, reporting any malfunction directly and receiving instructions for self-repair, where possible. The throughput is 190 samples per hour, and walk away mode and continuous loading are supported.

Westergren method

According to ICSH recommendations, ESR by the Westergren method was performed using a diluted sample and circular glass Westergren tube with an inner diameter of 2.55 mm. Briefly, the EDTA-anticoagulated blood was thoroughly mixed by complete inversion of the tube 20 times and diluted 4:1 using a trisodium citrate dihydrate solution (3.8%). The Westergren tube was allowed to stand for 60 min in an appropriate supporting rack in a stable, vertical position in a fume hood, at constant temperature (20–25 °C) and free from external influences such as vibrations, heat, and direct sunlight. The sedimentation rate was read by visual determination after 60 min as the distance from the top of the plasma level to the top of the Red Blood Cell (RBC) layer and recorded in mm/h. All Westergren analyses were performed by a single analyst to minimize pipetting and reading variations. The comparison study against the gold standard method was made on the same working day within 5 h of blood sampling.

Method comparison study

Samples of 132 females and 139 males (a total of 271) were selected among those over the age of 18 years [(median age 58, interquartile ranges (IQR): 42.25–70.75 years], spanning the entire ESR analytical range (from 2 to 120 mm), and with hematocrit value ≥ 0.35 L/L [0.41 ± 0.03 L/L, mean ± SD; 95% confidence interval (CI): 0.406–0.414). Hematocrit was determined as part of the complete blood count on Sysmex XN‐10 automated hematology analyzer (Sysmex Corporation, Kobe, Japan).

Undiluted EDTA samples were used first for the determination of ESR by VES-Matic 5, then the EDTA‐anticoagulated sample was diluted manually with 3.8% citrate solution in a ratio of 4:1, according to the ICSH recommendations for the Westergren method. Following the ICSH recommendations, method comparisons were further assessed in three subgroups, according to the ESR values obtained with the Westergren method, that is, low (<20 mm), middle (20–60 mm), and upper third (>60 mm) of the analytical range.

Precision study

Intra-run precision was estimated using 10 routine patient samples with ESR values covering the analytical range from 2 to 140 mm. All samples were analyzed on VES-Matic 5 for five replicates. Inter-run precision was performed using commercial quality control (QC) samples in the normal and abnormal range. Specifically, inter-run precision on VES-Matic 5 was assessed using two lots of the control material, the ESR Control Cube, composed of stabilized human blood at two levels for normal and abnormal values. For evaluation of the inter-run precision, controls were used three times a day on five consecutive days.

Sensitivity to fibrinogen

According to ICSH recommendations, the sensitivity of response to added fibrinogen was assessed in six healthy volunteers (3 females and 3 males) recruited from the laboratory staff. Human fibrinogen (Sigma-Aldrich, St Louis, MO, USA, expected concentration 20 g/L) was dissolved at 37 °C in 0.9% saline solution, and sterile-filtered using a 0.22 µm filter (Merck Millipore Ltd., County Cork, Ireland), as recommended by the manufacturer. Fibrinogen concentration obtained in stock solution was 14.8 g/L, measured using the automated Clauss method (Sysmex CS-5100, Sysmex Corporation, Kobe, Japan). Samples were prepared, with the addition of saline solution alone, or saline spiked with stock fibrinogen to reach concentrations of: 1.5, 3, 4.5, 6, and 8 g/L and added to each tube. ESR was determined first on the VES-Matic 5 analyzer and then by the original Westergren method. The baseline fibrinogen concentration (normal range 1.5–4.5 g/L) of six volunteers (shown as series in Figure 2), measured by the Clauss method, was the following: series 1: 1.9 g/L; series 2: 2.7 g/L; series 3: 2.7 g/L; series 4: 4.1 g/L; series 5: 2.4 g/L; series 6: 2.2 g/L. The sensitivity for fibrinogen in four of the volunteers was assessed to reach the concentration of 8 g/L.

Hemolysis interference

Twenty-eight routine patient samples, covering the whole ESR range, were selected for a hemolysis interference study. At first, ESR was determined by VES-Matic 5 analyzer and later hemolysis was induced in vitro by the addition of 50 µL of lysing solution (Osmored, Eurospital Spa, Trieste, Italy). To eliminate the dilution effect, 50 µL of blood was withdrawn from each tube before the addition of lysing solution. Hemolyzed samples were analyzed using the VES-Matic 5 and then by the Westergren method. Hemolysis index (HI) was determined on Architect c8000 (Abbott Laboratories, Abbott Park, IL, USA), which uses a spectrophotometric method incorporating four wavelength pairs and predefined constants to report numeric H-index into plasma-free hemoglobin results.

Triglycerides interference

TGs interference was evaluated in three routine patient samples, each with the following different values of ESR: 7, 32, and 138 mm/h, respectively. To evaluate interferences, a solution of TGs from a human plasma TG fraction (Lee Biosolutions, Maryland Heights, MO, USA, lot number W105417) was made, reaching a final concentration of 360 mg/dL (4.06 mmol/L) (Cobas Modular 8000, Roche Diagnostics S.p.A, Monza, Italy). Using the VES-Matic 5, ESR was evaluated first in each sample; then, aliquots of 150 μL of TG solution (0.61 mmol/L) were added three times, reaching the final addition of 450 μL of TG solution (1.83 mmol/L).

Anemia interference

Anemia interferences were evaluated in 141 routine patient samples, spanning the entire ESR analytical range (from 4 to 140 mm) and selected based on hemoglobin values (<109 g/L) and hematocrit values (<0.34 L/L) (Sysmex XN‐10, Sysmex Corporation, Kobe, Japan). ESR was first evaluated by the VES-Matic 5 analyzer and then using the Westergren reference method.

Verification of reference ranges

Reference ranges were verified according to the CLSI EP28-A3C document [7]. For transferability assessment ESR was determined both using VES-Matic 5 and the Westergren method in a total of 311 samples (163 healthy males and 148 healthy females, aged from 18 to 94 years). Samples were from routine outpatients that required a CBC, without the collection of additional test tubes for this study. Information such as food in the past 24 h, alcohol consumption, lifestyle, and health status was investigated at the time of collection. Results of the required laboratory tests were then verified. Samples were used for age and gender-related reference interval groups. Due to the difficulty of recruiting pediatric samples, only gender, and age-specific adult reference ranges were verified in this study, divided by age into three groups: from 18 to 50 years, 50–70 years, and older than 70 years.

The statistical significance was calculated at 95% of the reference interval. The following reference ranges were applied: under 50 years: 15 mm for males, and 20 for females; over 50 and below 70 years: <20 mm for males and <30 mm for females. Additional upper limits for people over 70 years old, were <30 mm for males and <35 for females. The ranges chosen for verification of the reference ranges obtained in this study are the most common given in the literature and the H02-A5 CLSI document [8], [9], [10].

Statistical analysis

The comparison of the VES-Matic 5 against the reference method was evaluated by calculating bias and limits of agreement using Bland–Altman analysis. Linear regression was carried out using Passing–Bablok regression to estimate constant and/or proportional difference, and Spearman’s rank correlation coefficient (ρ) was calculated to assess the strength and direction of the association between the compared data. Mean and median values, standard deviations (SD), and IQRs (25th and 75th quartiles) were also reported. For the precision study, coefficients of variation (CVs), mean values, SD, and confidence intervals (CI) at 95% were calculated. For the evaluation of interferences, a paired sample t-test and Wilcoxon signed-rank test were used for the comparison of groups. The D’Agostino–Pearson test was used to assess data distribution normality. One-way analysis of variance (ANOVA) test was used to evaluate interference studies. Statistical analysis was performed with MedCalc Statistical Software version 19.1 (MedCalc Software bv, Ostend, Belgium; https://www.medcalc.org; 2019). Precision and verification of reference ranges were performed with Analyse-it for Microsoft Excel software, release version 5.92 (Analyse-it Software, Ltd., Leeds, LS3 1HS, United Kingdom).

Results

Method comparison study

The mean ESR values obtained were 18.81 mm (95% CI: 16.27–21.36 mm) for the VES-Matic 5 and 18.25 mm (95% CI: 15.55–20.95 mm) for the gold standard Westergren method. The median ESR values were 11 mm (IQR: 5–23.75 mm) using VES-Matic 5 and 9 mm (IQR: 5–22.75 mm) with the Westergren method. The comparison of methods by Passing–Bablok analysis has shown a good agreement without systematic or proportional differences. The regression equation was y=−0.646 (95% CI: −1.00 to 0.11) + 0.979 (95% CI: 0.944–1.00)x (Figure 1A). The mean bias of −0.542 (95% CI: −1.06 to −0.01) was obtained by Bland–Altman analysis; the value of 8.03 for the upper limit and −9.11 for the lower limit can be considered acceptable limits, implicating a non-significant bias based on clinical criterion (Figure 1B). The Spearman’s coefficient of rank correlation (ρ) was 0.963, with its 95% CI from 0.953 to 0.971. The study comparison was also performed for the subgroups of results, based on the analytical interval, i.e., for the low (<20 mm), middle (20–60 mm), and upper range (>60 mm) of values and are presented in Table 1. Passing–Bablok regression analysis, Spearman’s rank correlation coefficient, and Bland–Altman analysis results were reported for each subgroup. For the low and middle range ESR values, the mean bias obtained by the Bland–Altman analysis can be considered acceptable from a clinical point of view.

Figure 1: 
Comparison between erythrocyte sedimentation rate values using the new VES-Matic 5 analyzer against the Westergren method as the gold standard (n=271).
(A) Passing–Bablok scatter diagram with an equation y=−0.646 (95% CI: −1.00 to 0.11) + 0.979 (95% CI: 0.944–1.00)x; (B) Bland–Altman scatter diagram with a mean bias of −0542 (95% CI: −1.06 to −0.01); the value of 8.03 for the upper limit and −9.11 for the lower limit can be considered acceptable limits, implicating a non-significant bias based on a clinical criterion.
Figure 1:

Comparison between erythrocyte sedimentation rate values using the new VES-Matic 5 analyzer against the Westergren method as the gold standard (n=271).

(A) Passing–Bablok scatter diagram with an equation y=−0.646 (95% CI: −1.00 to 0.11) + 0.979 (95% CI: 0.944–1.00)x; (B) Bland–Altman scatter diagram with a mean bias of −0542 (95% CI: −1.06 to −0.01); the value of 8.03 for the upper limit and −9.11 for the lower limit can be considered acceptable limits, implicating a non-significant bias based on a clinical criterion.

Table 1:

Comparison of the VES-Matic 5 with the reference Westergren method.

Equation Intercept (95% CI) Slope (95% CI) ρ (95% CI) Mean bias (95% CI)
Low (n=190) y=0.0 + 1.0x 0 (−0.33 to 0.0) 1.0 (1.0–1.11) 0.899 (0.868–0.923) −0.51 (−0.80 to −0.22)
Medium (n=61) y=5.16 + 0.92x 5.16 (2.0–8.0) 0.92 (0.80–1.0) 0.885 (0.815–0.930) −2.03 (−3.30 to −0.77)
High (n=20) y=27.7 + 0.60x 27.7 (15.8–36.49) 0.60 (0.49–0.78) 0.865 (0.685 to 0.946) 3.4 (−1.99 to 8.79)
  1. n, sample size; CI, confidence interval; ρ, Spearman’s rank correlation coefficient. According to ICSH recommendations results are reported for the low (<20 mm), middle (40–60 mm), and upper range (>60 mm) of the analytical interval.

Precision studies

In Table 2 the results of the intra-run precision study are presented, spanning a full range of values. ESR results are reported as mean value, 95% CI for the mean, SD, CV%, and the minimum and maximum values obtained. Five replicate measurements of the specimens showed a good level of precision that was below 10% (CV < 10%), except for two specimens, one with very low results, another one with a value of 11.2 mm, and a CV of 11.64%. In Table 3 the inter-run precision is reported. The evaluation was determined with two different lots of QC material, both in the normal and abnormal range, analyzed three times on five consecutive days. Inter-run precision was below 11% (CV 10.9%) for the normal level and <3% (CV 2.7%) for abnormal level. The Grubbs’ test did not reveal any outliers.

Table 2:

Intra-run precision of VES-Matic 5.

Samples n Mean, mm 95% CI SD, ± CV, % Minimum, mm Maximum, mm
1 5 3.2 2.16–4.24 0.45 26.15 2 4
2 5 6.8 6.24–7.35 0.45 6.60 6 7
3 5 11.2 9.58–12.82 1.30 11.64 9 12
4 5 16 15.12–16.88 0.77 4.42 15 17
5 5 31.8 30.76–32.84 0.84 2.63 31 33
6 5 44 39.88–48.12 3.32 7.53 41 49
7 5 52.8 49.58–56.01 2.59 4.90 49 56
8 5 91.4 88.28–94.52 2.51 2.75 87 93
9 5 97.8 93.46–102.14 3.49 3.57 92 101
10 5 138 132.45–143.55 4.47 3.24 130 140
  1. CI, confidence interval; SD, standard deviation; CV, coefficient of variation. ESR values are expressed as the mean, the 95% and SD of five determinations, whereas imprecision is expressed as the within-run coefficient of variation (CV%). The minimum and maximum ESR values obtained were reported.

Table 3:

Inter-run precision (CV%) of the VES-Matic 5 by analysis of three replicates of commercial control samples in the normal and abnormal range.

Control Lot Days Mean, mm SD, ± CV, %
Normal 720 5 6.5 0.7 10.9
Abnormal 720 5 37.2 1.0 2.7
  1. Erythrocyte sedimentation rate values are expressed as the mean, standard deviation (SD), and coefficient of variation (CV).

Sensitivity to fibrinogen

The sensitivity to increasing amounts of fibrinogen was determined and an excellent similar sensitivity was obtained both with the VES-Matic 5 and the Westergren reference method (Figure 2A and B). The assessment of the linearity of response between methods and the sensitivity was evaluated using the Bland–Altman statistical method, and no significant bias was observed (arithmetic mean: −0.7353; 95% CI from −1.547 to 0.0765; p=0.0743, NS). Calculation of the correlation coefficient (r=0.998, 95% CI from 0.996 to 0.999) and the slope (0.993, 95% CI from 0.971 to 1.014) gives an excellent assessment of response to fibrinogen of VES-Matic 5. The most significant rise of ESR related to increasing concentrations was observed for both methods at the final added fibrinogen concentration of 6 and 8 g/L. The highest ESR elevation was six and 9.47 times the baseline (mean value from 8.5 mm/h to 51.2 and 80.5 mm/h) for the Westergren method, while ESR elevation was 5.6 and 8.64 times the baseline (mean value from 9.3 mm/h to 51.67 and 80.3 mm/h) for the VES-Matic 5.

Figure 2: 
Sensitivity of ESR to added fibrinogen concentrations, measured with (A) the VES-Matic analyzer and (B) the Westergren method.
Figure 2:

Sensitivity of ESR to added fibrinogen concentrations, measured with (A) the VES-Matic analyzer and (B) the Westergren method.

Hemolysis interference

The results of the assessment of hemolysis interference were evaluated by the Bland–Altman method, and the mean of the differences was not significant (bias: 0.036, 95% CI from −22,896 to 2,3610). The results of the ANOVA test showed that the mean values of the subgroups did not differ significantly (p=0.972, NS) demonstrating that there are no hemolysis effects on ESR determination using VES-Matic 5. The median HI value was 5.2 g/L (ranging from 0.9 to 6.7 g/L of free hemoglobin).

Triglycerides interference

For the VES-Matic 5, the assessment of TGs interference on ESR determination was evaluated by ANOVA test, and the values obtained for the three samples did not differ significantly (p=1.00, not significant, NS), demonstrating that there are no significant effects on the ESR results due to the increase of TG concentration. The baseline values compared to those obtained in samples with the final concentration of 162 mg/dL of TGs were respectively: from 6 to 8 mm/h; from 41 to 30 mm/h; from 120 to 140 mm/h.

Anemia interference

To evaluate potential anemia interference, samples selected had hemoglobin values from 72 to 109 g/L (mean value: 96.82 ± 8.25 g/L, 95% CI from 95.72 to 97.93) and hematocrit values from 0.21 to 0.34 L/L (mean value: 0.297 ± 0.27 L/L, 95% CI from 0.29 to 0.30). The mean ESR value obtained by VES-Matic 5 was 57.11 ± 44.21 mm (95% CI: from 49.75 to 64.47 mm) while the mean ESR with the gold standard Westergren method was 49.53 ± 41.11 mm/h (95% CI: from 42.69 to 56.38 mm).

The regression equation y=1.80 (95% CI from −0.18 to 3.54) + 1.12 (95% CI from 1.06 to 1.18)x calculated by the Passing–Bablok regression analysis showed a good comparison of methods, with no statistically significant systematic differences while a minor constant and proportional difference was found, when comparing high values (Figure 3). The cumulative test (Cusum linearity test, p-value=0.17) indicated linearity between methods, while the Spearman’s coefficient of rank correlation (ρ) was 0.966 (95% CI: from 0.953 to 0.975).

Figure 3: 
Passing–Bablok scatter diagram when erythrocyte sedimentation rate values are obtained using the new VES-Matic 5 analyzer against the Westergren method as the gold standard. The regression equation y=1.80 + 1.12 showed a good comparison of methods, with no statistically significant systematic differences; for high values, there is a proportional difference that does not impact from a clinical point of view.
Figure 3:

Passing–Bablok scatter diagram when erythrocyte sedimentation rate values are obtained using the new VES-Matic 5 analyzer against the Westergren method as the gold standard. The regression equation y=1.80 + 1.12 showed a good comparison of methods, with no statistically significant systematic differences; for high values, there is a proportional difference that does not impact from a clinical point of view.

Reference range study

For each gender and age group, 95% of the reference upper limits obtained by the VES-Matic 5 were inside those reported in Materials and Methods section, therefore making it possible to confirm the reference limits of the present study. The upper limits obtained in this study are reported in Table 4.

Table 4:

Upper reference limit for the erythrocyte sedimentation rate obtained from the VES-Matic 5 and the Westergren method.

Gender and age, years n VES-Matic 5, mm Westergren, mm
Mean ± SD Upper limit Mean ± SD Upper limit
Male < 50 52 6 ± 6 15 (11–18)a 6 ± 6 14 (9–19)a
Male from 50 to 70 64 9 ± 7 19 (16–22)a 8 ± 7 18 (15–21)a
Male >70 49 17 ± 10 33 (30–37)a 14 ± 9 30 (24–33)a
Female <50 61 12 ± 8 25 (21–28)a 11 ± 8 22 (18–26)a
Females from 50 to 70 43 14 ± 8 27 (23–30)a 12 ± 7 24 (19–28)a
Female >70 44 19 ± 11 37 (33–41)a 16 ± 9 31 (26–35)a
  1. n, sample size; SD, standard Deviation. aBootstrap confidence interval (500 iterations) at 90%. Reference values are established locally in accordance with the CLSI EP28-A3C document (see text).

Discussion

ESR is still considered an overall test to assess the acute phase reactants, as a “sickness index” that clinicians have used for decision-making for over 50 years, in conjunction with the physical examination and clinical history of the patient [11, 12].

The diagnostic accuracy of ESR and C-reactive protein in acute inflammation diseases has recently been reassessed and has shown that ESR can provide valuable clinical information. For example, its determination may contribute to the management of COVID-19 patients in the determination of disease progression [13, 14]. ESR continues to demonstrate its unquestioned clinical usefulness in diagnosis, monitoring, and progression of response to therapy of certain diseases associated with immune response, such as rheumatoid arthritis or systemic lupus erythematosus [15], [16], [17]. In other arthritic disorders, ESR is one of the main laboratory tests in investigating patients presenting signs and symptoms of temporal arteritis or polymyalgia rheumatica [18, 19].

Automation of ESR testing, based on technological improvements in studies of the sedimentation phenomenon, has brought about the development of many automated instruments which are now available for routine use. It should be underlined that automation allows for workflow optimization, increases personnel safety by using closed systems, and shortens turnaround times. Consequently, updated recommendations published by ICSH addressed standardization and harmonization of the novel methods to decrease variation in the interpretation of ESR results [6].

In this study, following ICSH recommendations we evaluated the analytical performance of a new automated ESR analyzer, the VES-Matic 5, that uses the sedimentation principle and undiluted EDTA samples. This is an important issue because the most suitable specimen for ESR testing is that employed for CBC, i.e., samples anticoagulated with EDTA that avoid the collection of a second specimen if clinicians require both tests in the same patients. The use of undiluted or diluted specimens represents one of the most important testing variables. For automated ESR determination with diluted sodium citrate blood, test tubes are dedicated and the anticoagulated blood ratio is different from that of the citrate tube used for coagulation testing, which must never be used. To improve standardization, the anticoagulant type for specimens to use in ESR testing should be discussed again at an international level to achieve comparable results. The VES-Matic 5 comparison against the reference Westergren method, assessed in the diluted specimen, has shown excellent performance and agreement, with better data than other previous comparison studies [20, 21]. For ESR very low results, equal to 3.2 mm, the intra-run precision showed a CV% equal to 26.15; data should be judged on clinical interpretation because the minimum and maximum values vary from 2 to 4 mm. For other ranges of results, the imprecision was <10%, except for only one case with an ESR value of 11.2 mm, with a CV of 11.64%. In terms of inter-run precision, the system provided a satisfactory and acceptable performance that was better than those published in previous studies [22, 23]. As far as interferences are concerned, the VES-Matic 5 response to each level of fibrinogen concentration was in the agreement with Westergren, indicating that the new instrument is interchangeable with the gold standard method.

In hemolysis evaluation, the VES-Matic 5 did not show significant susceptibility to hemolysis, contrary to previous data observed using the instrument VES-Matic Cube 200, manufactured by the same company. For hemolysis interference we operated similarly to the study reported for the VES-Matic Cube 200, i.e., using spiking of native samples with a lysing solution. No significant decrease was observed for ESR values before and after lysis of the samples and no negative effects, similar to the studies on instruments that utilize photometric rheology principle, achieving harmonization in this issue [24, 25].

As compared to other studies, no significant decrease in ESR values was observed in TGs interference evaluation, but this study has some limitations because of the small number of samples assessed, and poor recovery of concentration of TGs [26], [27], [28].

Anemia can affect the erythrocytes sedimentation, causing an acceleration of the global phenomenon. We evaluated the effect of anemia by employing samples of 141 patients, with a hemoglobin level ranging from 72 to 109 g/L, minimizing variables that can affect the manual technique for the Westergren method. In this study, VES-Matic 5 showed a good comparison against Westergren by statistic Passing–Bablok regression and no significant systematic difference was found, except for ESR high values, when constant and proportional differences were found, clinically not significant. Our data are in agreement with the previous study [29].

ESR varies greatly with age and sex and can be influenced by lifestyle factors (physical activity, smoking, and alcohol consumption), and common metabolic abnormalities (obesity and related metabolic syndrome) [3031].

Although reference values should be established locally, many laboratories are adopting references published in papers or books or stated by the manufacturers [932, 33]. In this regard, some consideration should be made: reference ranges were published years ago when there were variations in methodology such as different pipettes and materials used in the Westergren method. In an effort, to achieve maximum diagnostic efficiency when interpreting the results of the ESR test, this issue deserves special attention.

In this study, VES-Matic 5 showed reference ranges that meet those obtained utilizing the Westergren method. The preliminary upper range of the present study should be interpreted considering undiluted samples and a non-parametric approach, statistical method recommended by CLSI. The highest ESR was found in healthy subjects over 70, different in men and women, and therefore separate values should be established for people in this age bracket.

This study assessed the analytical validity of the VES-Matic 5 analyzer and the comparability with the gold standard method. According to our findings, and considerating that there are useful commercial QC and suitable EQA materials and programs, the VES-Matic 5 can be employed appropriately for routine purposes. Finally, the VES-Matic 5 works using AI-based software for the reduction of pre-analytical errors including mislabeled, hemolyzed, or coagulated samples. Laboratory medicine has a central role in the diagnostic workup of many diseases and the application of AI can improve diagnostics through more accurate detection of pathology. Determination of erythrocyte sedimentation parameters utilizing AI has already been found to allow the monitoring of rare hereditary neurodegenerative diseases and further studies may reveal other possible clinical applications [34]. In conclusion, ESR is by no means an obsolete test.


Corresponding author: Prof. Mario Plebani, Department of Laboratory Medicine, University-Hospital of Padova, Unità Operativa Complessa di Medicina di Laboratorio, Azienda Ospedale-Università, Via Nicolò Giustiniani 2, 35128 Padova, Italy, Fax: 39-049-663240, E-mail:

Acknowledgments

The authors would like to thank Daniela Rinaldi and Monica Razzetti for the technical assistance and DIESSE Diagnostica Senese for kindly supplying reagents without in any way influencing the study design and data analysis.

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Informed consent was obtained from all individuals included in this study.

  5. Ethical approval: The local Institutional Review Board deemed the study exempt from review.

References

1. Grzybowski, A, Sak, J. Edmund Biernacki (1866–1911): discoverer of the erythrocyte sedimentation rate. On the 100th anniversary of his death. Clin Dermatol 2011;29:697–703. https://doi.org/10.1016/j.clindermatol.2011.08.033.Search in Google Scholar

2. Fabry, TL. Mechanism of erythrocyte aggregation and sedimentation. Blood 1987;70:1572–6. https://doi.org/10.1182/blood.v70.5.1572.bloodjournal7051572.Search in Google Scholar

3. Pribush, A, Meyerstein, D, Meyerstein, N. The mechanism of erythrocyte sedimentation. Part 1: channeling in sedimenting blood. Colloids Surf B Biointerfaces 2010;75:214–23. https://doi.org/10.1016/j.colsurfb.2009.08.036.Search in Google Scholar

4. Pribush, A, Meyerstein, D, Meyerstein, N. The mechanism of erythrocyte sedimentation. Part 2: the global collapse of settling erythrocyte network. Colloids Surf B Biointerfaces 2010;75:224–9. https://doi.org/10.1016/j.colsurfb.2009.08.037.Search in Google Scholar

5. Plebani, M. Erythrocyte sedimentation rate: innovative techniques for an obsolete test? Clin Chem Lab Med 2003;41:115–6. https://doi.org/10.1515/CCLM.2003.019.Search in Google Scholar

6. Kratz, A, Plebani, M, Peng, M, Lee, YK, McCafferty, R, Machin, SJ. International Council for Standardization in Haematology (ICSH). ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate. Int J Lab Hematol 2017;39:448–57. https://doi.org/10.1111/ijlh.12693.Search in Google Scholar

7. Clinical and Laboratory Standards Institute (CLSI). Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline, 3rd ed. CLSI Document EP28-A3C; 2010.Search in Google Scholar

8. Brigden, ML. Clinical utility of the erythrocyte sedimentation rate. Am Fam Physician 1999;60:1443–50.Search in Google Scholar

9. Böttiger, LE, Svedberg, CA. Normal erythrocyte sedimentation rate and age. Br Med J 1967;2:85–7.10.1136/bmj.2.5544.85Search in Google Scholar

10. Clinical and Laboratory Standards Institute (CLSI). Procedures for the erythrocyte sedimentation rate test. Approved standard, 5th ed. CLSI Document H02-A5; 2011.Search in Google Scholar

11. Bedell, SE, Bush, BT. Erythrocyte sedimentation rate. From folklore to facts. Am J Med 1985;78:1001–9. https://doi.org/10.1016/0002-9343(85)90224-4.Search in Google Scholar

12. Bray, C, Bell, LN, Liang, H, Haykal, R, Kaiksow, F, Mazza, JJ, et al.. Erythrocyte sedimentation rate and C-reactive protein measurements and their relevance in clinical medicine. Wis Med J 2016;115:317–21.Search in Google Scholar

13. Lapić, I, Padoan, A, Bozzato, D, Plebani, M. Erythrocyte sedimentation rate and C-reactive protein in acute inflammation. Am J Clin Pathol 2020;153:14–29. https://doi.org/10.1093/ajcp/aqz142.Search in Google Scholar PubMed

14. Lapić, I, Rogić, D, Plebani, M. Erythrocyte sedimentation rate is associated with severe coronavirus disease 2019 (COVID-19): a pooled analysis. Clin Chem Lab Med 2020;58:1146–8. https://doi.org/10.1515/cclm-2020-0620.Search in Google Scholar PubMed

15. Aletaha, D, Smolen, JS. Diagnosis and management of rheumatoid arthritis - a review. JAMA 2018;13:1360–72. https://doi.org/10.1001/jama.2018.13103.Search in Google Scholar

16. Smolen, JS, Aletaha, D, McInnes, IB. Rheumatoid arthritis. Lancet 2016;388:2023–38. https://doi.org/10.1016/s0140-6736(16)30173-8.Search in Google Scholar

17. Dima, A, Opris, D, Jurcut, C, Baicus, C. Is there still a place for erythrocyte sedimentation rate and C-reactive protein in systemic lupus erythematosus? Lupus 2016;25:1173–9. https://doi.org/10.1177/0961203316651742.Search in Google Scholar PubMed

18. Mahmood, SB, Nelson, E, Padniewski, J, Nasr, R. Polymyalgia rheumatica: an updated review. Cleve Clin J Med 2020;87:549–56. https://doi.org/10.3949/ccjm.87a.20008.Search in Google Scholar PubMed

19. Ling, ML, Yosar, J, Lee, BW, Shah, SA, Jiang, SW, Finniss, A, et al.. The diagnosis and management of temporal arteritis. Clin Exp Optom 2020;103:572–82. https://doi.org/10.1111/cxo.12975.Search in Google Scholar PubMed

20. Lapić, I, Piva, E, Spolaore, F, Tosato, F, Pelloso, M, Plebani, M. Automated measurement of the erythrocyte sedimentation rate: method validation and comparison. Clin Chem Lab Med 2019;57:1364–73. https://doi.org/10.1515/cclm-2019-0204.Search in Google Scholar PubMed

21. Lapić, I, Miloš, M, Tosato, F, Piva, E, Zadro, R, Rogic, D, et al.. Analytical validation of the iSED automated analyzer for erythrocyte sedimentation rate. Int J Lab Hematol 2020;42:109–15. https://doi.org/10.1111/ijlh.13120.Search in Google Scholar PubMed

22. Lapić, I, Piva, E, Spolaore, F, Musso, G, Tosato, F, Pelloso, M, et al.. Ves-Matic CUBE 200: is modified Westergren method for erythrocyte sedimentation rate a valid alternative to the gold standard? J Clin Pathol 2019;72:716–9. https://doi.org/10.1136/jclinpath-2019-205873.Search in Google Scholar PubMed

23. Schapkaitz, E, RabuRabu, S, Engelbrecht, M. Differences in erythrocyte sedimentation rates using a modified Westergren method and an alternate method. J Clin Lab Anal 2019;33:e22661. https://doi.org/10.1002/jcla.22661.Search in Google Scholar PubMed PubMed Central

24. Sezer, S, Yilmaz, FM, Kaya, O, Uysal, S. Evaluation of Ves-Matic Cube 200 for erythrocyte sedimentation rate determination. J Clin Lab Anal 2013;27:367–72. https://doi.org/10.1002/jcla.21612.Search in Google Scholar PubMed PubMed Central

25. Boğdaycioğlu, N, Yilmaz, FM, Sezer, S, Oğuz, E. Comparison of iSED and Ves-Matic Cube 200 erythrocyte sedimentation rate measurements with Westergren method. J Clin Lab Anal 2015;29:397–404. https://doi.org/10.1002/jcla.21786.Search in Google Scholar PubMed PubMed Central

26. Mahlangu, JN, Davids, M. Three-way comparison of methods for the measurement of the erythrocyte sedimentation rate. J Clin Lab Anal 2008;22:346–52. https://doi.org/10.1002/jcla.20267.Search in Google Scholar PubMed PubMed Central

27. Narang, V, Grover, S, Kang, AK, Garg, A, Sood, N. Comparative analysis of erythrocyte sedimentation rate measured by automated and manual methods in anaemic patients. J Lab Physicians 2020;12:239–43. https://doi.org/10.1055/s-0040-1721155.Search in Google Scholar PubMed PubMed Central

28. Hilder, FM, Gunz, FW. The effect of age on normal values of the Westergen sedimentation rate. J Clin Pathol 1964;17:292–3. https://doi.org/10.1136/jcp.17.3.292.Search in Google Scholar PubMed PubMed Central

29. Shearn, MA, Kang, IY. Effect of age and sex on the erythrocyte sedimentation rate. J Rheumatol 1986;13:297–8.Search in Google Scholar

30. Alende-Castro, V, Alonso-Sampedro, M, Vazquez-Temprano, N, Tunez, C, Rey, D, García-Iglesias, C, et al.. Factors influencing erythrocyte sedimentation rate in adults: new evidence for an old test. Medicine (Baltim) 2019;98:e16816. https://doi.org/10.1097/md.0000000000016816.Search in Google Scholar PubMed PubMed Central

31. Siemons, L, Ten Klooster, PM, Vonkeman, HE, van Riel, PL, Glas, CA, van de Laar, MA. How age and sex affect the erythrocyte sedimentation rate and C-reactive protein in early rheumatoid arthritis. BMC Muscoskel Disord 2014;15:368. https://doi.org/10.1186/1471-2474-15-368.Search in Google Scholar PubMed PubMed Central

32. Wetteland, P, Roger, M, Solberg, HE, Iversen, OH. Population-based erythrocyte sedimentation rates in 3910 subjectively healthy Norwegian adults. A statistical study based on men and women from the Oslo area. J Intern Med 1996;240:125–31. https://doi.org/10.1046/j.1365-2796.1996.30295851000.x.Search in Google Scholar PubMed

33. Caswell, M, Pike, LA, Bull, BS, Stuart, J. Effect of patient age on tests of the acute-phase response. Arch Pathol Lab Med 1993;117:906–10.Search in Google Scholar

34. Rabe, A, Kihm, A, Darras, A, Peikert, K, Simionato, G, Dasanna, AK, et al.. The erythrocyte sedimentation rate and its relation to cell shape and rigidity of red blood cells from chorea-acanthocytosis patients in an off-label treatment with dasatinib. Biomolecules 2021;11:727. https://doi.org/10.3390/biom11050727.Search in Google Scholar PubMed PubMed Central

Received: 2022-04-13
Accepted: 2022-04-21