Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a shared proactive risk analysis model

  • Viviane Van Hoof EMAIL logo , Suzanne Bench , Antonio Buño Soto , Peter P. Luppa , Anthony Malpass , Ulf Martin Schilling , Kevin D. Rooney , Adam Stretton and Andrei N. Tintu

Abstract

Objectives

Proposal of a risk analysis model to diminish negative impact on patient care by preanalytical errors in blood gas analysis (BGA).

Methods

Here we designed a Failure Mode and Effects Analysis (FMEA) risk assessment template for BGA, based on literature references and expertise of an international team of laboratory and clinical health care professionals.

Results

The FMEA identifies pre-analytical process steps, errors that may occur whilst performing BGA (potential failure mode), possible consequences (potential failure effect) and preventive/corrective actions (current controls). Probability of failure occurrence (OCC), severity of failure (SEV) and probability of failure detection (DET) are scored per potential failure mode. OCC and DET depend on test setting and patient population e.g., they differ in primary community health centres as compared to secondary community hospitals and third line university or specialized hospitals. OCC and DET also differ between stand-alone and networked instruments, manual and automated patient identification, and whether results are automatically transmitted to the patient’s electronic health record. The risk priority number (RPN = SEV × OCC × DET) can be applied to determine the sequence in which risks are addressed. RPN can be recalculated after implementing changes to decrease OCC and/or increase DET. Key performance indicators are also proposed to evaluate changes.

Conclusions

This FMEA model will help health care professionals manage and minimize the risk of preanalytical errors in BGA.

Introduction

Key reason for using point-of-care testing (POCT) is the rapid availability of results, allowing prompt clinical decision-making without the need to send samples to a central laboratory. For safe, effective, and person-centred care, it is imperative that POCT results are absolutely accurate and reliable. POCT technology is already well established in emergency departments (ED) and intensive care units (ICU). Although the use of this type of testing generally does not require specific technical laboratory skills, POCT provision and use should be guided by a clinical laboratory and performed by trained and certified personnel only [1].

The three phases of clinical laboratory testing: preanalytical, analytical and post-analytical also apply to POCT. This article focuses on the pre-analytical phase, known to be responsible for up to 62% of all errors in laboratory medicine [2, 3]. Some preanalytical risks apply to all laboratory tests, including POCT, e.g., wrong or absent sample identification, while other risks are specific for the central laboratory test or for the POCT under consideration.

POCT blood gas analysis (BGA) was chosen as subject for this risk analysis, as it is one of the most complex POCTs, combining the measurement of blood gases, electrolytes, haemoglobin, co-oximetry, and other parameters such as glucose, lactate, bilirubin, ionized calcium and ionized magnesium on one instrument.

BGA is recommended by the American Association for Respiratory Care and other cardiopulmonary care societies for evaluating a patient’s ventilatory, acid-base and/or oxygenation status, for evaluating a patient’s response to therapeutic interventions and for monitoring severity and progression of cardiopulmonary disease processes [4].

BGA is usually carried out in a POCT setting in a busy and stressful environment, such as the ED and the ICU, where time-critical clinical decisions are made in patient management. Several preanalytical aspects of POCT BGA are unique to this type of testing and the multiplicity of measured and calculated parameters amplifies the effects of potential preanalytical errors leading to wrong results that can have immediate negative impact on patient outcome.

Major risks arise from poor operator competency, lack of supervision, poor governance, failure to implement quality assurance processes, lack of understanding of the limitations of use and uncertainty on how to act on the results [5].

To address the growing concern surrounding preanalytical errors, we believe that there is need for a dedicated risk analysis template that unites the available literature, whilst offering a practical solution to respond to the requirements of the International Organization for Standardization (ISO) standards. ISO has developed quality systems to assess specific aspects of health services. The ISO 15189:2012 standard for clinical laboratories requires that the laboratory evaluates the impact of work processes and potential failures on examination results, as they affect patient care and safety, and that the laboratory modifies processes, to reduce or eliminate the identified risks, and documents decisions and actions taken [6]. The ISO 22870:2016 standard is specific for POCT and is based on ISO 15189:2012 [7]. As a result, ISO 15189:2012 requirements, including risk management, also apply to POCT. As described in these ISO documents, the POCT coordinator, designated by the laboratory, has a pivotal role in the organisation, realisation, and certification of user training as well as promoting awareness to users and care-givers about preanalytical errors in POCT and their severely negative impact on patient safety and care.

As appropriate tool we used FMEA (Failure Mode and Effect Analysis) to develop a proactive model that identifies preanalytical risks of POCT BGA, measures their potential impact on patient outcome, and can be used to monitor corrective actions, with the goal to diminish risks. FMEA is a well-known tool for the analysis of process failures in many fields, that is also applicable for use in healthcare [89]. The proposed model includes key performance indicators (KPIs) and we also briefly describe the use of Risk Ranking Tables. KPIs and Risk Ranking Tables are complementary or alternative ways to monitor the effect of corrective actions [10, 11]. To the best of our knowledge, there is no other literature proposing a specific risk analysis for POCT BGA.

Materials and methods

A multidisciplinary team, composed of laboratory and clinical staff from different European countries, actively involved in BGA, designed the FMEA template. It was built in a Microsoft Excel spreadsheet and was informed by a combination of focus group discussions involving the team members and data from published literature [12], [13], [14], [15], [16], [17], [18], [19], [20], [21], [22], [23], [24], [25], [26], [27], [28], [29], [30], [31], [32]. The FMEA risk analysis procedure is schematically represented in Figure 1. The process steps involved in the pre-analytical phase of BGA are listed in the proposed model, as well as the errors that can occur whilst performing these steps (potential failure mode), the possible consequences of these errors (potential failure effect), their potential causes, and examples of potential preventive and detection measures that can be undertaken to avoid the errors (current controls). In accordance with De Vries et al., a scale from one to four was proposed to score probability of failure occurrence (OCC) severity of failure (SEV) and probability of failure detection (DET) per potential failure mode [9] (Tables 1A1C). SEV, OCC, and DET are used to calculate the risk priority number (RPN = SEV × OCC × DET) [11]. The proposed model considers the classical blood gas parameters: arterial partial pressure of oxygen (PaO2), arterial partial pressure of carbon dioxide (PaCO2), pH, measured or calculated arterial oxygen saturation (SaO2), calculated bicarbonate (HCO3 ) and base excess (BE). These parameters can be reported on all blood gas analysers. The fraction of inspired oxygen (FiO2) and the patient’s body temperature can be introduced for calculation of PaO2/FiO2 and the alveolar-arterial gradient (A-a), and for temperature correction of pH and blood gas values respectively. Electrolytes such as sodium (Na+), potassium (K+) and chloride (Cl−) with calculated anion gap (AG), ionized calcium (Ca++), metabolites such as glucose (GLUC), lactate (LAC), bilirubin (BIL), and haematological parameters such as haemoglobin (Hb), haematocrit (Hct) and foetal haemoglobin (HbF) were also included as they can be measured by most blood gas analysers. Ionized magnesium (Mg++) is exceptionally measured and was also included, as were parameters measured by means of co-oximetry: oxygen saturation of haemoglobin (sO2), oxyhaemoglobin (O2Hb), deoxyhaemoglobin (HHb), carboxyhaemoglobin (COHb) and methaemoglobin (MetHb). Table 2 shows a snapshot of the FMEA model, containing the 12 process steps for BGA that were identified and one example per process step. Table 3 shows an example of a Risk Ranking Table calculated according to XFMEA [11]. In this example, the team performing the risk analysis decided 1° that a SEV = 1 does not need any corrective action, regardless of the value for OCC, 2° that a high severity (SEV = 4) will always give rise to a corrective action, whatsoever value for OCC, and 3° that a SEV of two or three will need corrective action only when there is a low probability of detection (DET ≥ 3). Table 4 shows the difference between monitoring the result of corrective actions after two revisions by means of the RPN value, with a user-defined threshold for RPN ≥ 10, or by means of a Risk Ranking Table using the same conditions as applied in Table 3.

Figure 1: 
Schematic representation of risk analysis by means of failure mode and effects analysis (FMEA), key performance indicators (KPI) and Risk Ranking Tables.
Figure 1:

Schematic representation of risk analysis by means of failure mode and effects analysis (FMEA), key performance indicators (KPI) and Risk Ranking Tables.

Table 1A:

Scoring table for severity of failure (SEV), according to De Vries et al. [9].

SEV score 1 2 3 4
Severity scale Minor event Moderate event Major event Catastrophic event
Patient outcome Neither injury nor increased length of stay or increased level of care Increased length of stay or increased level of care for one or two patients Permanent lessening of body functioning, disfigurement, surgical intervention required, increased length of stay of three or more patients Death or major permanent loss of function or suicide
Staff outcome First aid treatment only with no loss of time or restricted duty injuries or illness Medical expenses, lost time or restricted duty injuries or illness for one or two staff Hospitalization of one or two staff, or three or more staff experiencing lost time or restricted duty injuries or illnesses One death or hospitalization of three or more staff
Equipment outcome Damages < 10,000$ without adverse patient outcome 10,000$ ≤ damages < 100,000$ 100,000$ ≤ damages < 250,000$ Damages ≥ 250,000$
Table 1B:

Scoring table for probability of failure occurrence (OCC), according to De Vries et al. [9].

OCC score 1 2 3 4
Probability scale Remote Uncommon Occasional Frequent
Unlikely to occur; may happen sometime in 5–30 years Possible to occur; may happen sometime in 2–5 years Probably will occur; may happen several times in 1–2 years Likely to occur immediately or within a short period, may happen several times a year
Table 1C:

Scoring table for probability of failure detection (DET).

DET score 1 2 3 4
Detection scale Absolute certainty High Low Absolute uncertainty
Always detected Moderately high to high probability of detection (detected in ≥50% of cases) Moderately low to low probability of detection (detected in <50% of cases) Never detected
Table 2:

process steps of the failure mode and effects analysis (FMEA) POCT BGA, with examples of potential failure modes, current controls and key performance indicators (KPI) per process step.

Current controls
No. Process step Potential failure mode Potential failure effect SEV Potential causes OCC Examples of potential preventive measures Examples of potential detection measures DET RPN Definition of KPI KPI User defined criterium for KPI
1. Patient checks to be performed before starting the collection procedure
1.1. Check whether patient has aberrant hemoglobin, e.g. sickle cell anemia, fetal hemoglobin Erroneous procedure in this particular type of patient Inaccurate cooximetry readings. Falsely increased carboxyhaemoglobin. INCORRECT PATIENT DIAGNOSIS/TREATMENT 4 SEV refers to patient-related outcome Insufficient patient information. High oxygen affinity hemoglobins have altered dissociation curves and discrepant apparent oxygen saturation Training. Effective patient history taking/hand over. Perform measurement <15 min after blood collection if plastic syringe is used, or else use glass syringe Link with ICD code Percentage of samples from patients with the appropriate ICD code that were not measured in time/total number of samples
2. Entry of the blood gas request
2.1. Manual entry of the blood gas request on the instrument Missing tests or wrong tests performed on the instrument Need for repeat sample. 2 SEV refers to combined patient and staff related outcome Errors during manual entry of the requested tests on the instrument Electronic request system. Bidirectional communication between LIS and instrument
3. Identification of patient and user
3.1. Labelling of sample container (syringe) with patient identifying number or barcode before proceeding to blood collection
3.1.1. Manual identification of the sample
3.1.1.1. Identify patient and label sample manually No identification Syringe cannot be labeled. Need for repeat sample. 2 SEV refers to combined patient and staff related outcome Unidentified unconscious patient Dummy code Percentage of samples with dummy code/total number of samples
3.1.2. Barcode label printed after identification of the patient by reading patient’s barcode identifier on wrist band
3.1.2.1. Read barcode on wrist band and print label for sample No identification Syringe cannot be labeled 1 SEV refers to patient-related outcome No wrist band Manual identification with use of positive patient identifiers, name, date of birth, electronic ID via national number (e.g.,NHS number) Dummy code Percentage of samples with dummy code/total number of samples
3.2. Labelling syringe
3.2.1. Labelling syringe with request/sample ID Wrong identification Wrong label on sample, wrong (switched) blood gas results. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Barcode/label applied from other patient’s test request SOP, training Register misidentified samples Percentage of misidentified samples/total number of samples
3.3. Unique barcode on syringe
3.3.1. Read unique barcode on syringe and link to patient barcode No identification Blood gas results cannot be used. Need for repeat sample 2 SEV refers to combined patient and staff related outcome Barcode on syringe not linked to patient ID Warning by instrument, SOP, training Register unidentified samples Percentage of unidentified samples/total number of samples
3.4. Identification of the user
3.4.1. Staff verification of ID before injecting the sample (for audit trail) Lack of staff ID in procedure, result not traceable to user. Errors in process are not traceable to staff member, patient results go to wrong staff member, negligence or errors not detected 1 SEV refers to combined patient and staff related outcome Staff do not have ID badge and use/switch ID badges from/with other staff, or identify themselves on paperwork as other staff members. State registration, training, SOP Register samples with missing user identification. Warning for missing user identification by instrument Percentage of samples with missing user identification/total number of samples
4. Preparation of the patient
4.1. Allow time (at least 15–30 min) for stabilization following a change in position of the patient before collecting blood Patient not correctly prepared Wrong blood gas results. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Blood taken <15 min after change in position of the patient as postural changes affect ventilatory rate and depth Training
5. Circumstantial external factors
5.1. Disasters, pandemics, leading to a vast increase of ICU admissions Patients not correctly prepared, erroneous identification, erroneous measurements Wrong blood gas results. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Increased workload Disaster and pandemic training, effective multidisciplinary team communications, delegation of tasks
6. Patient characteristics
6.1. Remove traces of benzalkonium compounds (used as disinfectant) Incorrect procedure Invalid Na+ and/or Ca++. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample. Possible diagnostic delay with delay in therapeutic intervention. 4 SEV refers to combined patient and staff related outcome Insufficient patient information. Instrument flag for interference with benzalkonium compounds neglected by the user Training Instrument flag for interference with benzalkonium compounds Percentage of samples with instrument flag for interference with benzalkonium compounds/total number of samples
7. Infection risk
7.1. Check for infection in the patient when sampling blood Infected user No effect on patient results 4 SEV refers to staff related outcome Needle exposure Safe needles, training Register needle exposures Percentage of reported needle exposures/total number of samples
8. Type of sample
8.1. Select the appropriate syringe Wrong/inadequate syringe Wrong results. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Lack of training, distraction, stress. Inappropriate anticoagulants (other than heparin) such as oxalate and EDTA can interfere with electrolyte or enzymatic measurments by chelating divalent cations Staff training Register samples in wrong syringes Percentage of samples in wrong syringes/total number of samples
8.2. Arterial blood
8.2.1. Arterial puncture Erroneous procedure Wrong results. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Contamination with venous blood Training, skilled staff
8.3. Venous blood
8.3.1. Peripheral vein Wrong sample type if pO2 and pCO2 are clinically important in this patient Reduced pO2, possibly erroneous pCO2. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample. 4 SEV refers to combined patient and staff related outcome Large variability of the arterio-venous difference for pO2 Training
8.4. Capillary blood
8.4.1. Heel puncture Erroneous procedure Reduced pO2, possibly erroneous pCO2, increased K+. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample. 4 SEV refers to combined patient and staff related outcome Insufficient dilation of arterioles Training, skilled staff Comparison with results obtained in the central laboratory Percentage of samples with clinical significant difference between POCT and central laboratory results/total number of samples that were measured in both settings
9. Sample collection
9.1. Device storage Alteration of additive or syringe materials Altered additive stability, physical characteristics of the syringe leading to aberrant test results. INCORRECT PATIENT DIAGNOSIS/TREATMENT 4 SEV refers to patient-related outcome Device stored outside of manufacturer’s stated conditions Training, resilient product specification
10. Sample transport
10.1. Transport temperature Transport at 0 °C of plastic tube Increased or decreased pO2 (dependent on patient’s pO2), increased K+. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample 4 SEV refers to combined patient and staff related outcome Temperature too low for plastic tube: promotes movement of ambient air across plastic Training, logistics, point-of-care location of analyser
11. Choice of blood gas instrument
11.1. Injection of the sample Incorrect procedure Intra-patient “volatility” of blood gas and electrolyte results when samples of same patient are measured on different instruments 2 SEV refers to patient-related outcome Discrepancies between results of different blood gas instruments when the instruments are not properly managed SOPs should indicate which instrument(s) should be used for the ward and which instruments should be used as back-up. All instruments should be controlled and aligned. Differences between instruments should be kept as low as possible. If known differences cannot be avoided, the patient should be monitored with the same instrument Monitor differences between instruments by means of QC results Percentage of time instruments show clinically significant differences
12. Injection of the sample
12.1. Active injection of the sample (not on all instruments) Incorrect procedure Increased pO2 and K+. INCORRECT PATIENT DIAGNOSIS/TREATMENT. Need for repeat sample. Can cause damage to the instrument 4 SEV refers to combined patient, staff, and equipment and facility related outcome Excessive injection pressure (on particular instruments) Training Manufacturers of blood gas analyzers should develop methods to detect hemolysis in whole blood (currently hemolysis does not generate a specific flag for hemolysis on any instrument)
12.2. Active injection of the sample (not on all instruments) Incorrect procedure Infection risk for users. Can cause damage to the instrument 3 SEV refers to combined equipment and facility-related and staff-related outcome Blood spillage Training
  1. SEV, gravity of failure; OCC, probability of failure; DET, probability of failure detection; RPN, risk priority number.

Table 3:

Example of a risk ranking table constructed according to XFMEA methodology.

  1. For this example, the decisions taken by the team performing the risk analysis were: If SEV = 1, then corrective action is never required. If SEV = 4, then corrective action is always required. If 1 < SEV < 4, then corrective action is required if DET ≥ 3. SEV, Severity of failure; OCC, probability of failure occurrence. The numerical values in the cells represent probability of failure detection (DET).

Table 4:

Example of the use of a RPN Risk Ranking Table to monitor the effect of corrective actions, showing the difference between monitoring the effect of corrective actions by means of a Risk Ranking Table on one hand and by means of reduction in RPN on the other hand.

  1. For this example, the initial risk analysis and both revisions were performed by the team from Table 3. The cut-off for the RPN was set by the team at ≥10 (no corrective action needed when RPN < 10). After revision #1, corrective action is still needed, both according to the RPN and to the Risk Ranking Table (cfr. Table 3). After revision #2, corrective action is no longer needed according to the cut-off set for the RPN. After revision #2, corrective action is still needed according to the Risk Ranking Table, as SEV = 4. SEV, Gravity of failure; OCC, probability of failure; DET, probability of failure detection; RPN, Risk Priority Number.

Results

The FMEA model is provided as an open Excel template (Supplemental Material, File 1).

Users should decide which process steps, risks, and proposed controls for prevention and/or detection of the different failure modes apply to their setting and patient population and adapt the model accordingly. As SEV is relatively independent of settings and patient populations, a score for SEV is proposed. SEV scores can refer to purely patient-, staff-, or equipment/facility-related outcomes (Table 1A), or to a combination of two or all three of these categories. In the FMEA template the distinction is made by means of a colour scheme. A minority of SEV scores purely refers to patient outcomes. Most SEV scores refer to combined patient- and staff-related outcomes. Patient related outcomes are mostly increased test turnaround time (TAT), potentially wrong diagnosis with diagnostic delay resulting in a wrong or delayed therapeutic intervention, potential increase of the hospital length of stay (LOS) and temporary or permanent damage to the patient. Staff-related outcomes in these cases either relate to infection risk for the user or to an enhanced workload caused by the need for a repeat sample. Equipment or facility related outcomes relate to temporary or permanent instrument damage/instrument shutdown due to preanalytical errors. OCC and DET are dependent on settings and patient populations and should be scored by the performers of the risk evaluation to obtain a calculated RPN that is specific for their setting and patient population. The user should also determine the cut-off for the RPN that will trigger corrective actions, aiming at a reduction of OCC (preventive measures) and/or DET (detection measures). The % reduction of RPN after corrective actions is a measure of their success.

The template proposes several KPIs, that can also be used to monitor the effect of corrective actions. Data needed to calculate KPI’s can be retrieved from most lab information systems (LIS) or from the blood gas instrument. More complicated KPI’s including clinical data (e.g., ICD codes) need more advanced data-mining tools. Users should decide which KPI’s are feasible and applicable to their setting and patient population.

An additional way to manage corrective actions is the use of Risk Ranking Tables, wherein the initiation of corrective actions depends on user defined thresholds for OCC, SEV, and DET. The simple example of a Risk Ranking Table shown in Table 3 excludes corrective actions for errors with low SEV, low OCC and high probability of detection (low DET). When, over time, corrective actions have diminished the risk for errors with high SEV, high OCC and high DET, the user can lower the thresholds to also tackle errors with low SEV and/or low OCC and/or high probability of detection. The difference between monitoring the result of corrective actions by means of RPN or by means of a Risk Ranking Table is shown in Table 4. Based on RPN, corrective action would no longer be needed in this example after the second revision, as the RPN was lower than the threshold set by the user. However, according to the conditions set for the Risk Ranking Table, additional actions to lower DET and OCC would still be needed.

Discussion

Preanalytical POCT errors cause a considerable human, clinical and economic burden. This was recently demonstrated by Kazmierczak et al., who studied the impact of preanalytical POCT errors on productivity in a US ED [33]. The authors observed erroneous results in 6% of 15,479 i-STAT cartridges, of which 372 were unusable results. Errors for 163 out of 563 cartridges were definitely classified as originating from poor sample quality/improper sample handling. TAT and LOS were significantly longer with erroneous results. Direct costs over 2 years were 45,000 US$ and indirect cost was estimated between 486 and 729 h in avoidable nursing labor [33]. This study by Kazmierczak illustrates the need for proper risk management of the preanalytical phase in POCT.

FMEA is a generally accepted method for proactive risk evaluation, applicable in healthcare and more specific for clinical laboratory measurements [34], [35], [36], [37], [38]. It is important to emphasize that the FMEA model proposed in this article covers a broad range of work environments, as it was designed by a group of lab professionals and clinical staff from different European countries and with different backgrounds and profiles.

A recently learned lesson is that disasters, such as the COVID-19 pandemic, affect the way POCT is performed and increase the risk for pre-analytical errors. The vast number of ICU admissions caused by SARS-CoV-2 infections resulted in an increased workload with high and continuous stress. As a result of delocalization and conversion of non-ICU beds to ICU beds, ICU staff worked in non-familiar areas and non-ICU staff, who lacked training and experience, worked within an ICU setting. Off-line installation of extra blood gas instruments and open access of the instruments were additional risks. In these conditions it is almost certain that errors have been underreported. The presented FMEA covers these particular risks under “circumstantial external factors”, which is not per se a process step. To avoid the need for tracing the various causes of error due to the major risks associated with difficult circumstances, and to establish appropriate indicators and plan suitable strategies, users can choose to build a separate and more detailed FMEA for disasters departing from the presented model.

When an organization considers the FMEA type of risk evaluation, the analysis should be performed by an interdisciplinary team including lab manager(s), lab personnel, POCT coordinator(s), nurses, and other clinicians. The model that is presented here is constructed as a tool to manage risk, which can also be helpful when designing and implementing a new BGA network, but it is clearly not meant for ‘daily use’. Once the initial table has been filled in, the team should decide which risks are dealt with on a priority basis. Urgent issues need prompt action(s) and in this case the team must decide how soon the relevant part of the exercise has to be repeated to measure their effectiveness. In the absence of urgent issues, the full FMEA exercise can be repeated periodically (e.g., every two to three years) to evaluate the overall effect of the measures taken. Appropriate use of the template offers a practical solution to answer the requirements of the ISO standards, thereby answering its main aim, being the improvement of patient safety and patient care by diminishing the risks [6, 7].

While many of the pre-analytical steps in BGA are common to all laboratory tests, such as accurate specimen labelling, some are unique to this testing because of the physicochemical and biological properties of the analytes being measured. Biologic variation of some blood gas parameters is very low (e.g., pH, Na+) and even little error cannot be tolerated in order to interpret small but clinically relevant changes. Hence, the preanalytical steps must be perfectly followed and performed to ensure that the patient receives appropriate and timely therapy in response to correct analytical results.

For example, the preparation of the patient that requires a waiting time of 15–30 min after repositioning before taking a blood sample is often not applied [15]. As patient care is prioritised based on need, the risk of “skipping 15 min” is very high. Underlying reasons might be elucidated further but generally can be seen as a combination of departmental culture and (experienced) workload, insufficient knowledge and general stress, i.e. both internal and external factors. This can be addressed by both specific training regarding BGA and disaster training/stress management, and by quality control by local leadership. Interestingly, this error is more common when the ED or ICU is working with the normal flow or limited crowding, as moderate overcrowding will automatically provide the 15 min needed. As is shown with this example, but also throughout the whole template, the importance of training and ensuring best use of nurses’ time by streamlining preanalytical processes cannot be over-emphasised [39].

The importance of local factors was demonstrated by Auvet et al. [40]. By comparing BGA results for electrolytes and haemoglobin in a cardiac surgery operating room, a neurosurgical ICU and a polyvalent ICU, the authors show that identical analysers provided results of varying quality, depending on the local constraints of the ICUs. One of the most important findings of this study was that a stringent quality management can overcome these issues [40].

According to the guidelines of the American Association for Respiratory Care, the “gold standard” sample for BGA is arterial blood, collected by needle puncture of an artery or via an indwelling arterial catheter [4]. Capillary samples are not recommended to determine the oxygenation status of the patient, nor are central or peripheral venous samples recommended as a substitute for arterial blood measurement of pO2, pCO2, and pH. On the other hand, corrected central venous pH, pCO2 and bicarbonate have been shown to provide clinically accurate results and, for many patients, non-invasive pulse oximetry may be sufficiently accurate to determine the patient’s oxygenation status, avoiding the risks of an arterial puncture [1231]. Central or peripheral venous samples can be used for the fast measurement of electrolytes and metabolites, and to monitor the acid-base balance of patients (although severity of acidosis or alkalosis may be under- or overestimated) [12, 31].

A factor that may be overlooked is the importance of continuous in vitro cell metabolism for the accuracy of BG measurements. PaO2 for instance is affected by several factors, including the number of oxygen-consuming blood cells. The rate of in vitro oxygen consumption was found to be proportional to white-blood-cell count, platelet count and reticulocyte count [23]. Routine precautions, such as measuring the sample within 15 min (maximum 30 min), are not sufficient to prevent spurious hypoxemia due to significant in vitro oxygen consumption in the context of hyperleukocytosis (white-blood-cell count > 100 × 109/L) or extreme thrombocytosis (platelet count > 2000 × 109/L) [23]. Mature erythrocytes on the other hand lack mitochondria and contribute little to the total in vitro oxygen consumption, but they do metabolise glucose by anaerobic glycolysis and as a result, delayed measurement can cause spurious hypoglycaemia. Although the performance of small dedicated POCT devices for blood glucose measurement in critically ill patients has improved over time, BGA remains the golden standard for blood glucose measurement in these patients [25]. Incorrect POCT glucose measurement can be an important cause of falsely elevated glucose when using capillary blood, due to peripheral oedema. Treatment of these patients for hyperglycaemia can lead to serious hypoglycaemia, therefore arterial or venous blood samples are preferred for blood glucose measurement by means of blood gas analysers [25]. Brennan et al. reported that significant glucose contamination (3 mmol L−1 ± 3.4) was detected in all open arterial line systems up to an aspiration volume of five times the dead space, while no samples from the closed systems recorded glucose concentration >1 mmol L−1 [26]. The same authors also found that recommended minimal discard volumes are inadequate in the presence of glucose in the flush solution and can lead to high blood glucose readings, inappropriate insulin use, and iatrogenic neuroglycopenia. Closed-loop arterial sampling systems could be the universal solution [26].

Some substances that interfere with BGA measurements are mentioned in the template, e.g., salicylates and halogen ions, such as bromide, interfere with chloride measurement, while glycolic acid and D-lactate interfere with L-lactate measurement and benzalkonium interferes with electrolyte measurement [18, 20, 21]. Other molecules, such as ascorbic acid, bilirubin, citrate, EDTA, ethanol, heparin, glucose, paracetamol, salicylate and urea are also listed in BGA reference manuals as potentially interfering with lactate measurement.

In some instances, interchangeable use of electrolyte results (especially sodium) from direct ion-selective electrodes (dISE) – used in BGA – and indirect ion selective electrodes (iISE) – used in most routine central lab instruments – is not advisable, especially in a setting of hyperproteinaemia or hyperlipidaemia [32].

Several rating scales can be used to score SEV, OCC and DET (e.g., 1–4, 1–5, 1–10 etc.) but, as the scores are multiplied to obtain RPN, the same rating scale should be applied to SEV, OCC, and DET to avoid the appearance of skewing the resulting RPN [41]. In the present template a linear rating scale with consecutive numbers (1–4) was chosen. Alternatively, and as long as the same rating scale is applied to all three components, non-consecutive numbers can be chosen by the user (e.g., 1, 3, 5, 7), as they may allow more distinction between ratings and cause less debate amongst team members or, when the team wants to put more emphasis on the higher scores, a non-linear scoring scale can also be utilized (e.g., 1, 4, 9, 16) [41].

Only SEV is scored in the template, while OCC and DET are not, as they are highly dependent on the test setting and on the patient population and in any case, they should be scored by the multidisciplinary team. For example, OCC and DET will be different in a primary community health centre as compared to a secondary district general hospital and certainly as compared to a third line university hospital or specialized hospital e.g., trauma centre, cancer centre. OCC and DET will also differ whether the instruments are stand-alone or connected to a management software and a lab information system, whether patient identification is manual or automated and whether results are automatically transmitted to the patient’s electronic health record. When OCC and DET are scored for local settings and patient populations, the calculated RPN can be applied to determine the sequence and prioritise which risks need to be addressed. Therefore, the team should determine the cut-off for the RPN that triggers corrective action. After implementing changes to decrease the occurrence of errors, the RPN should be re-calculated to measure their positive effect [41]. Sometimes choices have to be made, as it is not always possible to reconcile different corrective actions e.g., although plastics are partially gas permeable as opposed to glass, glass syringes were largely replaced by plastic syringes due to safety concerns.

The exclusive use of RPN values for prioritisation of failures that need corrective actions should be considered with caution. As the scores for SEV, OCC, and DET are multiplied, small changes in one score can lead to large changes in RPN. For example, a failure mode with high DET, high OCC but low SEV would be prioritized the same as a failure mode with high DET, low OCC but huge SEV, despite having different risk implications [9, 42]. It may therefore not be appropriate to give equal weight to the three ratings that comprise the RPN. An organization may consider issues with high SEV and/or high OCC ratings to represent a higher risk than issues with high DET ratings. The analysis team may then decide to initiate a corrective action anyway because of the very high severity of the potential effect of the failure, even when the RPN is not high enough to trigger corrective action. In this regard the team may develop a Risk Ranking Table in addition to – or instead of – using RPN values for prioritization of the need for corrective actions [11]. Risk Ranking Tables identify whether corrective action is required based on the combination of individual values for SEV, OCC, and DET, thereby allowing for more nuance as compared to the use of the RPN value for evaluating the effect of corrective actions. Another way to monitor corrective actions is the use of KPIs [10]. KPIs are quantifiable measures used to evaluate whether objectives for performance are met. We present a number of KPI’s in our template. The team should decide which KPI’s are feasible and applicable to their setting and patient population. The team should also define criteria for these KPI’s, below which no corrective action is needed. The percentage change in the chosen KPI’s obtained after corrective actions is a measure for their effect. Introducing a priority scale for KPI’s could facilitate their gradual introduction into routine practice, by starting with a “mandatory” (score 1) and ending with a “valuable” (score 4) score, as was proposed by Plebani et al. for the reduction of preanalytical errors in the clinical laboratory [43].

Limitations of the study

The proposed template is based on a combination of data from published literature and the experience of the team members and their collaborators, but the template has not yet been clinically validated. Although we tried to be as complete as possible, not only including common errors but also rarely occurring causes of error, some potential hazards and unicorns may have been overlooked.

Conclusions

The proposed FMEA analysis model responds to the practical necessity of having a risk analysis tool that can be used to design and implement a BGA network and to monitor its improvement over time, according to the requirements for accreditation and certification. However, this is not the only practical consequence. In fact, we hope that this initiative may lay the foundations for a single FMEA model that is widely applicable in various organisational contexts at international level. This shared model for proactive risk analysis could become a starting point for quality comparisons between organisations, thanks to the sharing of the same monitoring indicators, as it happens with laboratory errors [43, 44]. This would make the effort to carry out such analyses even more useful and significant, as it could lead to an assessment of the organisation against the average international performance, if not against shared standards (benchmark).


Corresponding author: Viviane Van Hoof, MD, PhD, Prof. Em. Medical Biochemistry, Faculty of Medicine and Health Sciences, University of Antwerp, Campus Drie Eiken, Universiteitsplein 1, 2610 Wilrijk, Belgium, E-mail:

Acknowledgments

We are indebted to Becton, Dickinson and Company, Wokingham, UK  for logistic support of this project and for organizing the (virtual) meetings of the multidisciplinary international team.

  1. Research Funding: Not applicable.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Suzanne Bench is a National Institute for Health Research (NIHR) Senior Nurse and Midwife Research Leader. The views expressed in this article are those of the author(s) and not necessarily those of the NIHR, or the Department of Health and Social Care. All other authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Not applicable.

References

1. Institute of Biomedical Science. Point of care testing (Near-Patient testing) guidance on the involvement of the clinical laboratory. IBMS Professional Guidance. Available from: https://www.ibms.org/resources/documents/point-of-care-testing-near-patient-testing/ [Accessed 7 Mar 2022].Search in Google Scholar

2. Plebani, M, Sciacovelli, L, Aita, A, Pelloso, M, Chiozza, ML. Performance criteria and quality indicators for the pre-analytical phase. Clin Chem Lab Med 2015;53:943–8. https://doi.org/10.1515/cclm-2014-1124.Search in Google Scholar PubMed

3. Carraro, P, Plebani, M. Errors in a stat laboratory. Types and frequencies 10 years later. Clin Chem 2007;53:1338–42. https://doi.org/10.1373/clinchem.2007.088344.Search in Google Scholar PubMed

4. Davis, MD, Walsh, BK, Sittig, SE, Restrepo, RD. AARC Clinical practice guideline: blood gas analysis and hemoximetry:2013. Respir Care 2013;58:1694–703. https://doi.org/10.4187/respcare.02786.Search in Google Scholar PubMed

5. Orford, R. Policy on the management of point of care testing. What, when and how? Welsh Health Circular. Category Health Professional Letter. Issue date 12 July 2017. Available from: https://gov.wales/sites/default/files/publications/2019-07/policy-on-the-management-of-point-of-care-testing-poct.-what-when-and-how.pdf [Accessed 7 Mar 2022].Search in Google Scholar

6. ISO 15189:2012 medical laboratories – requirements for quality and competence. Available from: www.iso.org [Accessed 7 Mar 2022].Search in Google Scholar

7. ISO 22870:2016 point-of-care-testing – requirements for quality and competence. Intended to be used in conjunction with ISO 15189:2012. Available from: www.iso.org [Accessed 7 Mar 2022].Search in Google Scholar

8. Liu, H-C, Zhang, L-J, Ping, Y-J, Wang, L. Failure mode and effects analysis for proactive healthcare risk evaluation: a systematic literature review. J Eval Clin Pract 2020;26:1320–37. [Epub 2019 Dec 17]. https://doi.org/10.1111/jep.13317.Search in Google Scholar PubMed

9. De Vries, M, Fan, M, Tscheng, D, Hamilton, M, Trbovich, P. Clinical observations and a healthcare failure mode and effect analysis to identify vulnerabilities in the security and accounting of medications in Ontario hospitals: a study protocol. BMJ Open 2019;9:e027629. Available from: https://bmjopen.bmj.com/content/bmjopen/9/6/e027629.full.pdf [Accessed 7 Mar 2022]. https://doi.org/10.1136/bmjopen-2018-027629.Search in Google Scholar PubMed PubMed Central

10. Flegar-Mestric, Z, Perkow, S, Radeljak, A, Kardum Paro, MM, Prkacin, I, Devcic-Jeras, A. Risk analysis of the preanalytical process based on quality indicators data. Clin Chem Lab Med 2017;55:368–77. https://doi.org/10.1515/cclm-2016-0235.Search in Google Scholar PubMed

11. XFMEA for failure mode and effect analysis (FMEA). Examining risk priority numbers in FMEA. ReliaSoft. Hottinger Bruel & Kjaer inc. 2021. Available from: https://www.reliasoft.com/resources/resource-center/examining-risk-priority-numbers-in-fmea [Accessed 7 Mar 2022].Search in Google Scholar

12. Higgins, C. Central venous blood gas analysis. Available from: https://acutecaretesting.org/en/articles/central-venous-blood-gas-analysis [Accessed 7 Mar 2022].Search in Google Scholar

13. Higgins, C. Useful tips to avoid preanalytical errors in blood gas testing: pH, pCO2 and pO2 – 2016. Available from: https://acutecaretesting.org/en/articles/useful-tips-to-avoid-preanalytical-errors-in-blood-gas-testing-ph-pco2-and-po2 [Accessed 7 Mar 2022].Search in Google Scholar

14. Baird, G. Preanalytical considerations in blood gas analysis. Biochem Med 2013;23:19–27. https://doi.org/10.11613/bm.2013.005.Search in Google Scholar PubMed PubMed Central

15. Moran, RF, Feuilli, A. Critical care analytes: pre-analytical factors affecting result quality for combined blood gas and electrolyte systems. J Automat Chem 1989;11:201–5. https://doi.org/10.1155/s1463924689000404.Search in Google Scholar PubMed PubMed Central

16. Feyen, BFE, Coenen, D, Jorens, PG, Wouters, K, Maas, AIR, Van Hoof, V, et al.. Falsely elevated sodium levels during thiopental treatment in the ICU: technical interference on a laboratory device with important clinical relevance. Neurocritical Care 2013;18:64–9. https://doi.org/10.1007/s12028-012-9794-x.Search in Google Scholar PubMed

17. Dukic, L, Kopcinovic, LM, Dorotic, A, Barsic, I. Blood gas testing and related measurements: national recommendations on behalf of the Croatian society of medical biochemistry and laboratory medicine. Biochem Med 2016;26:318–36.10.11613/BM.2016.036Search in Google Scholar PubMed PubMed Central

18. Oyaert, M, Van Maerken, T, Bridts, S, Van Loon, S, Laverge, L, Stove, V. Analytical and pre-analytical performance characteristics of a novel cartridge-type blood gas analyzer for point-of-care and laboratory testing. Clin Biochem 2018;53:116–26. https://doi.org/10.1016/j.clinbiochem.2018.01.007.Search in Google Scholar PubMed

19. Lippi, G, Fontana, R, Avanzini, P, Sandei, F, Ippolito, L. Influence of spurious hemolysis on blood gas analysis. Clin Chem Lab Med 2013;51:1651–4. https://doi.org/10.1515/cclm-2012-0802.Search in Google Scholar PubMed

20. Vasuyattakul, S, Lertpattanasuwan, N, Vareesangthip, K, Nimmannit, S, Nilwarangkur, S. A negative anion gap as a clue to diagnose bromide intoxication. Nephron 1995;69:311–3. PMID: 7753265. https://doi.org/10.1159/000188476.Search in Google Scholar PubMed

21. Kaul, V, Imam, SH, Gambhir, HS, Sangha, A, Nandavaram, S. Negative anion gap metabolic acidosis in salicylate overdose--a zebra. Am J Emerg Med 2013;31:1536.e3–4. [Epub 2013 Jul 16]. https://doi.org/10.1016/j.ajem.2013.05.031.Search in Google Scholar PubMed

22. Muhm, JM. Predicted arterial oxygenation at commercial aircraft cabin altitudes. Aviat Space Environ Med 2004;75:905–12.Search in Google Scholar

23. Hess, C, Nicholls, A, Hunt, WB, Suratt, PM. Pseudohypoxemia secondary to leukemia and thrombocytosis. N Engl J Med 1979;301:361–3. https://doi.org/10.1056/nejm197908163010706.Search in Google Scholar PubMed

24. Knowles, TP, Mullin, RA, Hunter, JA, Douce, FH. Effects of syringe material, sample storage time and temperature on blood gas and oxygen saturation in arterialized human blood samples. Respir Care 2006;51:732–6.Search in Google Scholar

25. Eerdekens, G-J, Rex, S, Mesotten, D. Accuracy of blood glucose measurement and blood glucose targets. J Diabetes Sci Technol 2020;14:553–9. https://doi.org/10.1177/1932296820905581.Search in Google Scholar PubMed PubMed Central

26. Brennan, KA, Eapen, G, Turnbull, D. Reducing the risk of fatal and disabling hypoglycaemia: a comparison of arterial blood sampling systems. Br J Anaesth 2010;104:446–51. https://doi.org/10.1093/bja/aeq027.Search in Google Scholar PubMed

27. Astles, JR, Lubarsky, D, Loun, B, Sedor, FA, Toffaletti, JG. Pneumatic transport exacerbates interference of room air contamination in blood gas samples. Arch Pathol Lab Med 1996;120:642–7.Search in Google Scholar

28. Jorens, PG, Demey, HE, Schepens, PJC, Coucke, V, Verpooten, GA, Couttenye, MM, et al.. Unusual D-lactic acid acidosis from propylene glycol metabolism in overdose. J Toxicol Clin Toxicol 2004;42:163–9. https://doi.org/10.1081/clt-120030942.Search in Google Scholar PubMed

29. CLSI. Blood gas and pH analysis and related measurements; approved guideline – 2nd ed. CLSI Document C46-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2009. Available from: www.clsi.org [Accessed 9 Sep 2021].Search in Google Scholar

30. The Point of Care Testing Committee of the IFCC. A primer of point-of-care blood gas testing for laboratorians; Nov 2018. Available from: https://www.ifcc.org/media/477840/2018-08-blood-gas-document-final-nov_2018.pdf [Accessed 7 Mar 2022].Search in Google Scholar

31. Treger, R, Pirouz, S, Kamangar, N, Corry, D. Agreement between central venous and arterial blood gas measurements in the intensive care unit. Clin J Am Soc Nephrol 2010;5:390–4. https://doi.org/10.2215/CJN.00330109.Search in Google Scholar PubMed PubMed Central

32. Chopra, P. Kumar Datta S. Discrepancies in electrolyte measurement by direct and indirect ion selective electrodes due to interferences by proteins and lipids. J Lab Phys 2020;12:84–91. https://doi.org/10.1055/s-0040-1713690.Search in Google Scholar PubMed PubMed Central

33. Kazmierczak, SC, Morosyuk, S, Rajkumar, R. Evaluation of preanalytical point-of-care testing errors and their impact on productivity in the emergency department in the United States. J Appl Lab Med 2022;7:650–60. https://doi.org/10.1093/jalm/jfab158.Search in Google Scholar PubMed

34. Capunzo, M, Cavallo, P, Boccia, G, Brunetti, L, Pizzuti, S. A FMEA clinical laboratory case study: how to make problems and improvements measurable. Clin Leader Manag Rev 2004;18:37–41.Search in Google Scholar

35. Southard, PB, Kumar, S, Southard, CA. A modified delphi methodology to conduct a failure modes effects analysis: a patient-centric effort in a clinical medical laboratory. Qual Manag Health Care 2011;20:131–51. https://doi.org/10.1097/qmh.0b013e318213b079.Search in Google Scholar

36. Jiang, Y, Jiang, H, Ding, S, Liu, Q. Application of failure mode and effects analysis in a clinical chemistry laboratory. Clin Chim Acta 2015;48:80–5. https://doi.org/10.1016/j.cca.2015.06.016.Search in Google Scholar PubMed

37. Serafini, A, Troiano, G, Franceschini, E, Calzoni, P, Nante, N, Scapellato, C. Use of a systematic risk analysis method (FMECA) to improve quality in a clinical laboratory procedure. Ann Ig Med Preventiva Comunita 2016;28:288–95. https://doi.org/10.7416/ai.2016.2108.Search in Google Scholar PubMed

38. Mendes, ME, Ebner, PAR, Romano, P, Pacheco Neto, M, Sant’anna, A, Sumita, NM. Practical aspects of the use of FMEA tool in clinical laboratory risk management. J Bras Patol Med Lab 2013;49:174–81. https://doi.org/10.1590/s1676-24442013000300004.Search in Google Scholar

39. Bench, S. Critical care nurses’ views and experiences of preanalytical factors influencing point-of-care testing A qualitative study. Point Care 2020;19:97–100. https://doi.org/10.1097/POC.0000000000000212.Search in Google Scholar

40. Auvet, A, Espitalier, F, Grammatico-Guillon, L, Nay, M-A, Elaroussi, D, Laffon, M, et al.. Preanalytical conditions of point-of-care testing in the intensive care unit are decisive for analysis reliability. Ann Intensive Care 2016;6:57. https://doi.org/10.1186/s13613-016-0152-6.Search in Google Scholar PubMed PubMed Central

41. Manufacturing technology committee – risk management working group risk management training guides. Failure Modes and Effects Analysis Guide. Draft proposal; 2008. Available from: https://pqri.org/wp-content/uploads/2015/08/pdf/FMEA_Training_Guide.pdf [Accessed 7 Mar 2022].Search in Google Scholar

42. Dean Franklin, B, Shebl, NA, Barber, N. Failure mode and effects analysis: too little for too much? BMJ Qual Saf 2012;21:607–11. https://doi.org/10.1136/bmjqs-2011-000723.Search in Google Scholar PubMed

43. Plebani, M, Sciacovelli, L, Aita, A, Chiozza, ML. Harmonization of pre-analytical quality indicators. Biochem Med 2014;24:105–13. https://doi.org/10.11613/bm.2014.012.Search in Google Scholar

44. Sciacovelli, L, Lippi, G, Sumarac, Z, West, J, Garcia del Pino Castro, I, Furtado Vieira, K, et al.. Quality indicators in laboratory medicine: the status of the progress of IFCC working group “Laboratory errors and patient safety” project. Clin Chem Lab Med 2017;55:348–57. https://doi.org/10.1515/cclm-2016-0929.Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0319).


Received: 2021-09-29
Accepted: 2022-05-05
Published Online: 2022-05-24
Published in Print: 2022-07-26

© 2022 Viviane Van Hoof et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.