The simple reproducibility of a measurement result does not equal its overall measurement uncertainty

  • Mauro Panteghini EMAIL logo

To the Editor,

In reading the recently published discussion paper by Coskun et al. [1], I was happy to see that many of the concepts we previously promoted for correctly estimating measurement uncertainty (MU) in medical laboratories were endorsed [2, 3]. The “top-down” approach estimating MU of laboratory results is now officially recognized by the ISO Technical Specification 20914 that provides a practical guidance to be applied in medical laboratory settings for estimating MU of values produced by procedures intended to measure biological measurands [4]. It seems, however, that Authors did not fully realize that the inspiring concept behind this approach relies on the definition of MU across the entire traceability chain, starting with the uncertainty of reference materials (uref), extending through the IVD manufacturers and their processes for assignment of calibrator values and uncertainty (ucal), and ending with the random variability of measuring systems (uRw). Although correctly, the authors only discuss how to derive this last component of MU by using internal quality control (IQC) data, ignoring, however, that the MU estimate must include all uncertainties introduced by the selected calibration hierarchy for the measurand, beginning with the highest available reference down to the assigned value of the calibrator for the commercial IVD medical device. Although reference material providers, IVD manufacturers, and medical laboratories have different roles and independent tasks across the metrological traceability chain, their performances contribute together to the MU of patient results [5], [6], [7]. The crucial point is that the estimated MU must be always combined at each level of the employed traceability chain. Particularly, the MU at the level of clinical samples (uresult) must be the combination of all uncertainty contributions accumulated across the entire traceability chain. This refutes the common misconception, which also shine through the proposal by Coskun et al., that the simple reproducibility of a measurement result equals its overall MU. A correct estimate of MU of laboratory results is indeed not possible without ucal, which in turn should include uref. In the European market, the information about ucal shall be provided on request to the professional end-users. Sometimes, calibrators are offered without MU, but it is up to the laboratory professionals to ask manufacturers and obtain this information for the correct estimate of uresult [8].

The characteristics of the IQC material to be used for uRw estimate have been also previously described in detail and should be carefully considered [2, 3, 5, 9]. The material should be different from that used to check the correct alignment of the measuring system, be commutable and with concentration(s) corresponding to the decision cut-point(s) employed in the medical application of the test. Therefore, suggesting that “calculating MU for normal level IQC material” is a rational compromise does not consider the third requirement. IQC materials for estimating uRw should have analyte concentration levels close to clinical decision limits or, at least, to employed reference limits. This is important because, for most, if not all laboratory tests, MU varies with the analyte concentration, usually decreasing with increasing concentrations.


Corresponding author: Mauro Panteghini, Research Centre for Metrological Traceability in Laboratory Medicine (CIRME), Università degli Studi di Milano, Milan, Italy, E-mail:

  1. Research funding: None declared.

  2. Author contributions: Single author contribution.

  3. Competing interests: Author states no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Not applicable.

References

1. Coskun, A, Theodorsson, E, Oosterhuis, WP, Sandberg, S. Measurement uncertainty for practical use. Clin Chim Acta 2022;531:352–60. https://doi.org/10.1016/j.cca.2022.04.1003.Search in Google Scholar PubMed

2. Braga, F, Panteghini, M. The utility of measurement uncertainty in medical laboratories. Clin Chem Lab Med 2020;58:1407–13. https://doi.org/10.1515/cclm-2019-1336.Search in Google Scholar PubMed

3. Braga, F, Pasqualetti, S, Aloisio, E, Panteghini, M. The internal quality control in the traceability era. Clin Chem Lab Med 2021;59:291–300. https://doi.org/10.1515/cclm-2020-0371.Search in Google Scholar PubMed

4. ISO/TS 20914:2019. Medical laboratories – practical guidance for the estimation of measurement uncertainty, 1st ed. Geneva, Switzerland: ISO; 2019.Search in Google Scholar

5. Braga, F, Infusino, I, Panteghini, M. Performance criteria for combined uncertainty budget in the implementation of metrological traceability. Clin Chem Lab Med 2015;53:905–12. https://doi.org/10.1515/cclm-2014-1240.Search in Google Scholar PubMed

6. Infusino, I, Panteghini, M. Measurement uncertainty: friend or foe? Clin Biochem 2018;57:3–6. https://doi.org/10.1016/j.clinbiochem.2018.01.025.Search in Google Scholar PubMed

7. Braga, F, Panteghini, M. Defining permissible limits for the combined uncertainty budget in the implementation of metrological traceability. Clin Biochem 2018;57:7–11. https://doi.org/10.1016/j.clinbiochem.2018.03.007.Search in Google Scholar PubMed

8. Braga, F, Panteghini, M. Verification of in vitro medical diagnostics (IVD) metrological traceability: responsibilities and strategies. Clin Chim Acta 2014;432:55–61. https://doi.org/10.1016/j.cca.2013.11.022.Search in Google Scholar PubMed

9. Braga, F, Infusino, I, Panteghini, M. Role and responsibilities of laboratory medicine specialists in the verification of metrological traceability of in vitro medical diagnostics. J Med Biochem 2015;34:282–7. https://doi.org/10.1515/jomb-2015-0004.Search in Google Scholar PubMed PubMed Central

Received: 2022-06-27
Accepted: 2022-07-01
Published Online: 2022-07-11
Published in Print: 2022-09-27

© 2022 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Downloaded on 29.3.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2022-0618/html
Scroll to top button