Licensed Unlicensed Requires Authentication Published by De Gruyter September 15, 2022

Total serum vitamin B12 (cobalamin) LC-MS/MS assay as an arbiter of clinically discordant immunoassay results

  • Ruiping Zhang , Xiaoli Ma , Yutong Zou , Ling Qiu , Danchen Wang , Yueming Tang , Yongtong Cao , Songlin Yu EMAIL logo and Xinqi Cheng EMAIL logo

Abstract

Objectives

Measurement of the serum levels of vitamin B12 (VB12) is key for evaluating VB12 deficiency-dependent anemia. Immunoassay, the major method for determining VB12, tends to give false-normal results because of the presence of anti-intrinsic factor (IF-Ab) or other factors such as heterophilic antibodies et al. This study aimed to develop a liquid chromatography tandem mass spectrometry (LC-MS/MS) method that is helpful for distinguish false normal VB12 results measured by the immunoassay.

Methods

Different forms of VB12 were derivatized into CN-B12, which was collected through solid-phase extraction and analyzed via LC-MS/MS. 236 serum samples were measured both by LC-MS/MS and immunoassay, results were compared, and the IF-Ab effect was evaluated.

Results

The LC-MS/MS assay afforded a linear slope from 20 to 4,000 pmol/L for CN-B12. OH-VB12, methyl-VB12, and CoA-VB12 showed recovery within 89.3–109.5%. The intra-assay CV of VB12 was 2.6–4.1%, whereas the total CV was 9.3–9.8%. Passing–Bablok regression between LC-MS/MS and immunoassay results showed that the slope was 1.085 and the intercept was −15.691. The Bland–Altman plot showed that the mean difference and difference% were −34.6 pmol/L and 0.3%, respectively. Inter-rater agreement analysis showed that the linear weighted kappa value was 0.885, implying good agreement between the two methods. However, two samples were falsely elevated and one sample was falsely normal in the immunoassay compared with LC-MS/MS. The LC-MS/MS method helped in the distinction of false-normal VB12 results shown by the immunoassay.

Conclusions

The VB12 LC-MS/MS method can be used as an arbiter of clinically discordant immunoassay results.


Corresponding authors: Songlin Yu, MD and Xinqi Cheng, MD, Department of Clinical Laboratory, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, No. 1 Shuaifu Yuan, Dongcheng District, Beijing 100730, P.R. China, Phone: +86 01069159707, Fax +86 01069159712, E-mail: (S. Yu), (X. Cheng)

Ruiping Zhang, Xiaoli Ma and Yutong Zou contributed equally to this article.


Funding source: National Key Research and Development Program of China http://dx.doi.org/10.13039/501100012166

Award Identifier / Grant number: 2021YFC2401100

Funding source: National High Level Hospital Clinical Research Funding

Award Identifier / Grant number: 2022-PUMCH-A-138

  1. Research funding: This work was funded by the National Key Research and Development Program of China (No. 2021YFC2401100), and National High Level Hospital Clinical Research Funding (2022-PUMCH-A-138). The funding bodies were not involved in conducting or analyzing this work or in the decision to publish. The authors have no additional financial disclosures.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Informed consent was obtained from all individuals included in this study.

  5. Ethical approval: Research involving human subjects complied with all relevant national regulations, institutional policies and is in accordance with the tenets of the Helsinki Declaration (as revised in 2013), and has been approved by the authors’ Institutional Review Board (Peking Union Medical College Hospital) or equivalent committee. (ZS-2795).

References

1. Carmel, R. Biomarkers of cobalamin (vitamin B-12) status in the epidemiologic setting: a critical overview of context, applications, and performance characteristics of cobalamin, methylmalonic acid, and holotranscobalamin II. Am J Clin Nutr 2011;94:348S–58S. https://doi.org/10.3945/ajcn.111.013441.Search in Google Scholar PubMed PubMed Central

2. Wainwright, P, Narayanan, S, Cook, P. False-normal vitamin B12 results in a patient with pernicious anaemia. Clin Biochem 2015;48:1366–7. https://doi.org/10.1016/j.clinbiochem.2015.08.009.Search in Google Scholar PubMed

3. Carmel, R, Agrawal, YP. Failures of cobalamin assays in pernicious anemia. N Engl J Med 2012;367:385–6. https://doi.org/10.1056/NEJMc1204070.Search in Google Scholar PubMed

4. Yang, DT, Cook, RJ. Spurious elevations of vitamin B12 with pernicious anemia. N Engl J Med 2012;366:1742–3. https://doi.org/10.1056/NEJMc1201655.Search in Google Scholar PubMed

5. Hamilton, MS, Blackmore, S, Lee, A. Possible cause of false normal B-12 assays. BMJ 2006;333:654–5. https://doi.org/10.1136/bmj.333.7569.654-c.Search in Google Scholar PubMed PubMed Central

6. Soleimani, R, Favresse, J, Roy, T, Gruson, D, Fillee, C. Macro vitamin B12: an underestimated threat. Clin Chem Lab Med 2020;58:408–15. https://doi.org/10.1515/cclm-2019-0999.Search in Google Scholar PubMed

7. Bowen, RA, Drake, SK, Vanjani, R, Huey, ED, Grafman, J, Horne, MK3rd. Markedly increased vitamin B12 concentrations attributable to IgG-IgM-vitamin B12 immune complexes. Clin Chem 2006;52:2107–14. https://doi.org/10.1373/clinchem.2006.073882.Search in Google Scholar PubMed

8. Szterk, A, Roszko, M, Malek, K, Czerwonka, M, Waszkiewicz-Robak, B. Application of the SPE reversed phase HPLC/MS technique to determine vitamin B12 bio-active forms in beef. Meat Sci 2012;91:408–13. https://doi.org/10.1016/j.meatsci.2012.02.023.Search in Google Scholar PubMed

9. D’Ulivo, L, Yang, L, Ding, J, Pagliano, E, Leek, DM, Thibeault, MP, et al.. Determination of cyanocobalamin by isotope dilution LC-MS/MS. Anal Chim Acta 2017;990:103–9. https://doi.org/10.1016/j.aca.2017.07.065.Search in Google Scholar PubMed

10. CLSI. Liquid chromatography-mass spectrometry methods; Approved Guideline. CLSI document C62-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.Search in Google Scholar

11. CLSI. CLSI document EP15-A2. User verification of performance for precision and trueness ag, 23d. Wayne: CLSI; 2005.Search in Google Scholar

12. Yadav, M, Trivedi, V, Upadhyay, V, Shah, G, Baxi, GA, Goswami, S, et al.. Comparison of extraction procedures for assessment of matrix effect for selective and reliable determination of atazanavir in human plasma by LC-ESI-MS/MS. J Chromatogr, B: Anal Technol Biomed Life Sci 2012;885–886:138–49. https://doi.org/10.1016/j.jchromb.2011.12.031.Search in Google Scholar PubMed

13. Ma, X, Zou, Y, Tang, Y, Wang, D, Zhou, W, Yu, S, et al.. High-throughput analysis of total homocysteine and methylmalonic acid with the efficiency to separate succinic acid in serum and urine via liquid chromatography tandem mass spectrometry. J Chromatogr B Anal Technol Biomed Life Sci 2022;1193:123135. https://doi.org/10.1016/j.jchromb.2022.123135.Search in Google Scholar PubMed

14. Ricos, C, Alvarez, V, Cava, F, Garcia-Lario, JV, Hernandez, A, Jimenez, CV, et al.. Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation; 2014. Available from: http://www.westgard.com/biodatabase1.htm [Accessed 14 Mar 2022].Search in Google Scholar

15. MedCalc. https://www.medcalc.org/manual/kappa.php [Accessed 5 Mar 2022].Search in Google Scholar

16. Hadinata Lie, A, Chandra-Hioe, MV, Arcot, J. Sorbitol enhances the physicochemical stability of B12 vitamers. Int J Vitam Nutr Res 2020;90:439–47. https://doi.org/10.1024/0300-9831/a000578.Search in Google Scholar PubMed

17. Merrigan, SD, Yang, DT, Straseski, JA. Intrinsic factor blocking antibody interference is not detected in five automated cobalamin immunoassays. Am J Clin Pathol 2014;141:702–5. https://doi.org/10.1309/AJCPPWHSB94FTOGU.Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0523).


Received: 2022-05-30
Accepted: 2022-08-30
Published Online: 2022-09-15
Published in Print: 2023-01-27

© 2022 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 2.5.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2022-0523/html