Remote HbA1c testing via microsampling: fit for purpose?

  • Nick Verougstraete , Veronique Stove and Christophe P. Stove ORCID logo EMAIL logo

Abstract

The collection of capillary blood microsamples via finger-prick has several advantages over traditional blood collection. It is considered convenient and more patient-centric, enabling collection of the sample by the patient at her/his home with subsequent analysis in the lab following postal shipment. Determination of the diabetes biomarker HbA1c in self-collected microsamples to remotely monitor diabetes patients seems to be a very promising option which could eventually lead to better treatment adaptations and disease control. This is especially convenient/relevant for patients living in areas where venipuncture is impractical, or to support virtual consultations using telemedicine. Over the years, a substantial numbers of reports on HbA1c and microsampling have been published. However, the heterogeneity of the applied study designs and data evaluation is remarkable. This review provides a general and critical overview of these papers, along with specific points of attention that should be dealt with when aiming at implementing microsampling for reliable HbA1c determination. We focus on the used (dried) blood microsampling techniques, collection conditions, stability of the microsamples, sample extraction, analytical methods, method validation, correlation studies with conventional venous blood samples and patient satisfaction. Lastly, the possibility of using liquid instead of dried blood microsamples is discussed. Liquid blood microsampling is expected to have similar advantages as dried blood microsampling and several studies suggest it to be a suitable approach to collect samples remotely for subsequent HbA1c analysis in the lab.

Introduction

Hemoglobin A1c (HbA1c) is a glycated protein which is formed in vivo by the non-enzymatic reaction of glucose with the N-terminal valine residue of beta globin chains of HbA. Determination of HbA1c in blood is used for the diagnosis and monitoring of type 1 and 2 diabetes mellitus as its concentration is a reflection of the average glycemic control over the past two to three months, considering an average life span of normal red blood cells of 120 days. In comparison to blood glucose determinations, HbA1c is not subject to wide day-to-day and hour-to-hour fluctuations [1]. HbA1c values are expressed as a proportion of total hemoglobin concentration in % (National Glycohemoglobin Standardization Program (NGSP) units) or mmol/mol (System International (SI) units used by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) network). An HbA1c value of ≥6.5 % or 48 mmol/mol is used as a threshold for the diagnosis of diabetes [2]. Pre-diabetes is defined by HbA1c values between 5.7 and 6.4 % (39–47 mmol/mol) [3]. Alternatively, diabetes can be diagnosed by measuring fasting plasma glucose concentrations or by performing a two-hour plasma glucose tolerance test [1]. HbA1c results have a strong predictive value for the occurrence of diabetes complications. Therefore, during treatment, HbA1c levels should remain below 7 % (53 mmol/mol) to prevent the development and progression of microvascular long-term complications (retinopathy, neuropathy and kidney disease). Monitoring the glycemic status is recommended at least twice per year in patients with stable glycemic control or every three months in patients not meeting glycemic treatment goals or with recent treatment changes [4]. In conditions associated with altered red blood cell turnover (for example hemolytic anemia and recent blood loss or transfusion) other parameters should be used to diagnose or monitor diabetes.

HbA1c should be measured by a method standardized to the IFCC reference measurement procedure [5] and preferably certified by the NGSP [6]. The presence of hemoglobin variants in blood samples may interfere with HbA1c testing, leading to incorrect results and consequently inappropriate treatment decisions. For each HbA1c analytical method used, the user should be aware of the impact of the presence of a hemoglobin variant and/or elevated fetal hemoglobin (HbF) [7, 8].

HbA1c levels are traditionally/usually measured in whole blood obtained via an invasive venous blood sampling procedure, performed by qualified medical personnel. However, different (dried) blood microsampling techniques can be used, which offer several advantages over traditional blood collection [9]. A blood microsample refers to a sample which is typically less than 50 µL and is mostly taken by a finger-prick using a safety lancet, representing a ‘minimally invasive’ sample collection. This, combined with the lower sample volume, renders this type of sampling particularly relevant for use in pediatric populations. Moreover, the simplicity of the sampling procedure implies that collection of (dried) microsamples does not require a phlebotomist in a medical center, but can be performed by the patients themselves, in their home environment. In addition, logistics for transportation and storage can be simplified by using dried blood microsamples [10]. After collection, and drying if working with dried samples, the samples can be sent via standard postal services to the clinical laboratory for analysis. It can be arranged that patients send their samples a few days before their consultation. In this way, the actual HbA1c value of the patient will be available at the time of the patient’s visit, allowing insight into the patient’s most recent glycemic status. As a result, better adaptation of the treatment and glycemic control is to be expected [11]. Another option to have the actual HbA1c result readily available at the time of the patients’ clinical consultation, is the use of point of care testing (POCT) [12]. However, POCT is not within the scope of this review and will not be discussed further. In addition, microsampling is also a valuable tool in population-based epidemiology studies [13], [14], [15]. Obtaining blood samples via a finger-prick may increase the availability of HbA1c testing in areas where venipuncture is impractical [16]. Moreover, mailing samples by postal service can be advantageous for patients living in rural and remote areas with limited access to phlebotomy centers [17, 18]. Lastly, drying of a sample reduces biohazard risks that are associated with liquid whole blood samples.

Despite the general health threat and associated social distancing rules, the recent COVID-19 pandemic also resulted in positive effects on the healthcare landscape, triggering and speeding up many reforms, aimed at coping with issues like patient follow-up and at improving access to regular healthcare. The shift to virtual consultation for follow-up of chronic conditions using telemedicine is one of these reforms and can now be further implemented beyond the pandemic. In support of telemedicine, remote collection of (micro)samples has shown to be a valuable and attractive alternative to traditional sampling [19]. Besides the use of dried blood spots (DBS) for SARS-CoV-2 serologic testing [20], also programs to monitor diabetes patients with HbA1c testing on DBS were set up [21], [22], [23]. In this way, accessibility to testing results and accompanying optimal diabetes care were guaranteed [23]. The feasibility of remote HbA1c testing via DBS to monitor the glycemic status of patients treated via telemedicine during the COVID pandemic was demonstrated by Roberts et al.: upon request of a diabetes provider, DBS kits were mailed to the patients. After collection of the DBS at home, the samples were sent back to the lab and HbA1c results were provided prior to the telemedicine visit. However, only 53.4 % of the patients returned their DBS to the lab and a substantial number of these DBS samples was of insufficient quality [21], suggesting that there is room for improvement.

While the pandemic may have been a facilitator of home sampling studies, the application of dried blood microsampling strategies for measuring HbA1c stems from long before the pandemic, with a substantial number of papers having been published on this topic. However, the heterogeneity of the applied study designs and data evaluation is remarkable. The goal of this review is to present a general overview of the multitude of reported methods (shown in Table 1) and to discuss/address specific points of attention that should be dealt with when aiming at implementing dried or liquid blood microsampling for reliable HbA1c monitoring. Whereas virtually any methodology may be capable of delivering ‘a’ result, the key question that should be asked is: does application of a given methodology under real-life conditions result in a correct and reliable result? Unfortunately, this is a question that is too often not addressed.

Table 1:

Overview of published dried blood microsample HbA1c-based methods.

Reference Microsampling technique Complete DBS vs. punch (P) (diameter) Extraction/elution solution Analytical technique Application + included patients/samples Sample collection Time between collection and analysis Specific remarks
Egier et al. [13] DBS P 3/16 inch 1 mL hemolyzing buffer Ion-exchange HPLC (Bio-Rad Variant II) Comparison between (lab-prepared) venous DBS and whole blood from 85 clinical specimens Whole blood applied on filter paper in the lab Within 96 h of sample collection
Jeppson et al. [24] DBS P 3 mm 75 µL Na phosphate-citrate Triton X-100 buffer + cysteine-containing buffer Ion-exchange HPLC (in house) Comparison between capillary DBS and venous blood from 41 patients Capillary collection by participants under supervision at outpatient diabetes unit NA
Lacher et al. [14] DBS P 3.2 mm Buffer Ion-exchange HPLC (Bio-Rad Variant II) Population-based epidemiology study with 386 DBS samples Capillary collection by experienced phlebotomists in controlled environment Refrigerated up to 7 days
Maleska et al. [26] DBS P 6 mm 400 µL DW buffer Ion-exchange HPLC (Arkray ADAMS HA-8180V) Comparison between (lab-prepared) venous DBS and whole blood samples from 217 participants 75 µL whole blood spotted on filter paper NA Full analytical validation under ISO15189 guideline
Pollock et al. [25] DBS ? ? Ion-exchange HPLC (Tosoh G7) NA Blood spotted on filter paper Up to 6 days Focus on HbA1c as marker in newborn screening
Thomas et al. [15] DBS ? ? Ion-exchange HPLC (Bio-Rad D-10) Comparison between venous DBS and venous blood from 143 subjects Venous blood was spotted on filter papers both in controlled clinic settings and in field-type setting ? Evaluation of the influence of environmental conditions during DBS preparation
Crawford et al. [34] DBS P 91 mm2 2000 µL hemolyzing reagent Immuno-turbidimetry (Roche Tina-quant) Comparison between HbA1c measured from dried blood fractions of collection device to venous whole blood samples from healthy donors Both assisted- as self-collected capillary blood samples via finger-prick NA Samples were collected via a device which uses laminar-flow migration to separate cellular components from plasma. HbA1c was measured on the dried RBC portion (and cholesterol on the dried plasma).
Elliott et al. [18] DBS P 6 mm 500 µL hemolyzing reagent Immuno-turbidimetry (Roche Tina-quant) Comparison between capillary DBS and venous whole blood from 50 diabetes patients Capillary collection under supervision and by participants at home DBS included in study up to 7 days Comparison between DBS collected under supervision at the lab and unsupervised at participants’ home
Fokkema et al. [28] DBS P 3 mm 250 µL hemolyzing buffer Immuno-turbidimetry (Roche Tina-quant) Comparison between capillary DBS and venous whole blood from 93 diabetes patients Capillary collection by laboratory personnel and by participants at home Within 10 days Comparison between DBS collected by laboratory personnel and by participants at home
Hu et al. [30] DBS P 3.2 mm 300 µL hemolyzing reagent Immuno-turbidimetry (Olympus AU HbA1c) NA ? NA
Jones et al. [16] DBS 2 P 1/8 inch 500 µL hemolyzing reagent Immuno-turbidimetry (Roche Tina-quant) Comparison between (lab-prepared) venous DBS and whole blood from 73 routine HbA1c samples 20 µL whole blood spotted on filter paper NA
Lakshmy et al. [29] DBS P 6 mm 400 µL hemolysis reagent Immuno-turbidimetry (Agappe latex method) Comparison between (lab-prepared) venous DBS and whole blood samples from 30 diabetes patients 10 µL whole blood spotted on filter paper 15 days when stored at 4 °C
Mastronardi et al. [31] DBS P 3.2 mm Haemolysing reagent Immuno-turbidimetry (Thermo Fisher HbA1c) Comparison between (capillary and venous) DBS and whole blood from 115 participants Capillary collection by participants in a controlled research setting Up to 14 days, results were acceptable up to 7 days Application of correction equations taken into account time between sample collection and analysis
Miller et al. [32] DBS P 3 mm Deionized water Immuno-turbidimetry (Kamiya Biomedical kit) Comparison between capillary DBS and venous blood from 125 DBS DBS were collected at the participants’ home by health technicians or field interviewers ? Performance of collecting DBS by experienced health workers vs. unexperienced interviewers was compared
Parkes et al. [27] DBS ? ? Immuno-turbidimetry (Roche Tina-quant) 1625 DBS from 59 diabetes patients Capillary collection by patient in-clinic setting 3–12 days after sample collection
Scherer et al. [33] DBS ? ? Immuno-turbidimetry (Thermo Fisher HbA1c) 14,243 DBS cards from 2,237 participants at risk of developing diabetes DBS either taken by a nurse at the clinical site or by the participants at home Range from 0 to 400 days Multicenter study focused on the impact of different storage conditions
Henderson et al. [35] DBS P 3.2 mm 100 mM ammonium bicarbonate + trifluoroethanol LC-PRM-MS (in house) Comparison between venous whole blood and DBS obtained from 36 healthy volunteers Capillary DBS were obtained from each volunteer by a trained phlebotomist NA Simultaneously measurement of HbA1c, apoA-I and apoB in DBS
Verougstraete et al. [9] DBS + VAMS P 3 mm DBS: 250 µL deionized water; VAMS: 400 µL deionized water Ion-exchange HPLC (Tosoh G8) Capillary DBS and VAMS were compared to liquid blood results from 86 diabetes patients (40 adults and 44 children) 1 DBS and 2 VAMS were collected by the patients at their home Up to 6 days Collection of two types of dried blood microsamples: DBS and VAMS
Verougstraete et al. [38] Wet VAMS NA 800 µL deionized water Ion-exchange HPLC (Tosoh G8) Comparison between wet VAMS and liquid capillary blood from 76 pediatric diabetes patients 2 VAMS samples were collected by the patients at home and tips were brought into water within 30min after collection Exclusion criteria: up to 6 days Evaluation of adapted sampling protocol by collecting VAMS tip in water
Lima et al. [39] VAMS NA 20 mM ammonium bicarbonate + deoxycholic acid LC-MS/MS (in house) NA Capillary VAMS were collected via fingerprick from volunteers NA
Hall et al. [37] HemaSpot™ HemaSpot™ filter blade One blade in 1 mL hemolysis/wash solution Ion-exchange HPLC (Tosoh G8) Comparison between capillary DBS and capillary liquid blood from 40 insulin requiring adult patients Capillary collection by a study researcher Up to 5 days Evaluation of the impact of time between HemaSpot™ DBS preparation and HbA1c analysis
Hall et al. [11] HemaSpot™ HemaSpot™ filter blade One blade in 1 mL hemolysis/wash solution Ion-exchange HPLC (Tosoh G8) Comparison between venous blood and capillary DBS collected at clinic and at home from 128 diabetes patients Capillary collection by participants under guidance of study nurse and subsequently unsupervised at home Within 4 days of preparation Comparison between HemaSpot™ DBS collected at home and supervised at the clinic. Conclusions are not sufficiently supported by data.
Acar et al. [40] Bloodstains NA Dissolved in water Ion-exchange HPLC (Tosoh G8) NA 50 µL whole blood pipetted on glass and fabric NA Forensic investigation
  1. Papers are ordered per microsampling technique, per analytical technique and lastly, in alphabetical order of the first author. Papers of special interest/relevance have been highlighted in bold.

Used microsampling technique and conditions of sampling

To date, most of the papers on dried blood microsampling have used DBS sampling [9, 13], [14], [15], [16, 18, 24], [25], [26], [27], [28], [29], [30], [31], [32], [33], [34], [35]. After pricking the finger, multiple blood drops are collected on a dedicated piece of filter paper (typically Whatman 903 or PE226). Subsequently, the samples are dried for 2–3 h, after which they are stored in sealed plastic bags together with a desiccant. In a typical DBS workflow, a 3–6 mm diameter disc is punched from the DBS, followed by extraction of this subpunch. Already in 1982 Goldstein et al. reported on the measurement of glycated protein in whole blood deposited on filter papers [36]. Also variations on the conventional filter paper cards have been used. In two reports by Hall and colleagues, DBS were collected using HemaSpot™ devices. These devices comprise a plastic cover enclosing eight wedge-shaped filter papers (or “blades”) with a built-in desiccant, to promote drying of the samples. The blood flows through a central opening in the device to fill all the blades simultaneously [11, 37]. Also the determination of HbA1c in samples collected via volumetric absorptive microsampling (VAMS) has been reported [9, 38, 39]. VAMS is a dried blood based sampling technique in which a fixed volume of blood is collected by an absorbent tip attached to a plastic handler. Collection of VAMS samples has been reported to be more straightforward compared to DBS collection [9]. Moreover, HbA1c has also been determined in dried red blood cell (RBC) fractions obtained via a collection device using laminar-flow migration [34] and in bloodstains, both on cotton fabric and glass surfaces [40].

Besides the used microsampling technique itself, it is also important to look at the conditions in which the samples were collected in the different studies, as this can have an important influence on the quality of the dried blood sample and subsequently on the obtained HbA1c result. Not only the context of the sampling (clinical setting vs. home environment), but also whether the samples were collected by an experienced phlebotomist or by the patient her/himself should be taken into account. Prior to routine implementation, the clinical acceptability of the dried blood method should be evaluated. To this end, it is important that during method validation study samples are collected in a context reflecting the future routine use. For example, when aiming at home-sampling, HbA1c results of samples collected at home by study participants must be compared with the conventional results. It is not sufficient to perform an evaluation on DBS that were artificially generated in a lab setting by spotting venous whole blood onto filter paper [13, 15, 16, 25, 26, 29, 30], as the latter approach does not cover issues that may be encountered during e.g. transport of samples at ambient temperature. In some studies, capillary DBS were collected by trained phlebotomists instead of by the patients [14, 35, 37] or participants collected capillary samples supervised in a controlled research setting [24, 31].

Several studies comparing HbA1c results obtained from capillary DBS collected by experienced phlebotomists or supervised in a controlled environment with those obtained from DBS collected by the patients themselves at their home did not find a bias between both sample types [11, 18, 28]. However, using a similar set-up, others did report significant differences, although without apparent cause (time between sample preparation and analysis was excluded as a potential source) [11]. In a study conducted by Miller and colleagues, the performance of collecting DBS samples by experienced health technicians and unexperienced field interviewers at the participant’s home was compared. DBS samples, collected by both groups, were of similar good quality and the obtained HbA1c DBS results were not significantly different, indicating that DBS can be collected by unexperienced non-phlebotomists in a non-clinical environment [32]. In a substantial number of studies the dried blood microsamples were collected unsupervised by the participants at their home [9, 11, 18, 28, 33, 38]. This requires a thorough explanation concerning the sampling technique. To this end, apart from the instructions given during inclusion, a sampling brochure could be added to the home sampling kit [9, 38] or a link to a brief instruction video could be provided (see for example a video which was used in a large-scale alcohol biomarker study [41]).

Stability, transport and storage

As dried blood microsamples are intended to be collected by patients in their home environment, the samples need to be transported to the analyzing lab after collection. This implies that it should be evaluated whether the time between collection and analysis and the shipment conditions may have an influence on the HbA1c result. In the following section the stability of HbA1c in dried blood microsamples will be discussed. The impact of time between collection and analysis is also further implemented in the discussion on correlation experiments.

In general, drying of a blood sample is associated with increased stability of the analyte. For measuring HbA1c in routine settings the added value of this potentially increased stability is somewhat limited, given the stability of HbA1c in conventional liquid whole blood samples. Liquid whole blood for HbA1c analyses should be stored in the refrigerator (4 °C) no longer than 7 days and for long-term storage, samples should be stored at −70 °C or lower [42, 43]. However, when considering sample transportation and storage, working with dried blood microsamples does offer logistical advantages. E.g., dried microsamples can be transported via regular mail (under ambient conditions) and stored for prolonged periods at room temperature. This is especially relevant in low-resource settings [44]. Transport and storage of microsamples also requires less space [14]. Although most studies do not report issues regarding instability during transport, the time between collection and analysis does seem to have an impact (see part on correlation studies further). For long-term storage, DBS should be stored at −70 °C or less, similar as for liquid whole blood samples [13, 24, 42].

If specific weather conditions are to be expected during sample transport, this should be evaluated during method validation. Samples mailed via standard postal services are typically prone to temperature and humidity variations. For example, elevated temperatures are to be expected in a sun-exposed mailbox during summer.

The results from stability experiments performed in various studies are summarized in Table 2, with quite striking differences in outcome. Several explanations have been proposed for the apparent discrepancies between studies. It is hypothesized that HbA1c may be formed by ex vivo glycation of hemoglobin due to the contact between hemoglobin and glucose, i.e. the ongoing Amadori arrangement of the labile A1c form. This reaction is believed to be time- and temperature-dependent, so varying temperatures during shipment of the samples would add more variability to the dried blood HbA1c results [11, 18, 28, 33, 39]. A protocol was introduced in which filter papers for generating DBS were pretreated with a citrate-based solution in order to prevent possible ex vivo glycation of hemoglobin in the blood spot. However, pretreatment of filter papers did not improve assay results. This finding should be interpreted with caution as the authors did not present data/results for their experiment [30].

Table 2:

Results from stability experiments of HbA1c in different microsamples.

Reference Microsample Storage condition Observed effect on HbA1c values
Acar et al. [40] Bloodstains 18–22 °C and relative humidity of 50 % Values decreased over time in both fabric (after 28 days) and glass surface (after 56 days)
Fokkema et al. [28] DBS RT Increase of values from 0 to 10 days (irrespective whether exposed to air or stored in sealed plastic bag)
Jones et al. [16] DBS RT; 4 °C; 25 °C and 40 °C Stable up to 44 days storage at RT and up to 9 days at 4, 25 and 40 °C
Lakshmy et al. [29] DBS 4 °C No significant differences between results after 10 and 15 days compared to day 0 values
Maleska et al. [26] DBS RT Significant decrease after 7 days of storage
Parkes et al. [27] DBS 35–65 °C Stable for 12 h at 55 °C; values decreased at higher temperatures with largest decrease for DBS kept at 65 °C for 24 h
Thomas et al. [15] DBS RT and 4–6 °C (in darkness) Stable for 10 days in all conditions
Verougstraete et al. [38] Wet VAMS Tips in water (within 30min after collection) at ambient temperature Stable for 6 days
  1. Papers are displayed in alphabetical order of the first author. RT, room temperature.

Lastly, several studies provided insight into the average time interval between sample collection and arrival in the lab: samples were transported via standard postal services and arrived on average between 3 and 7 days after the collection [9, 16, 18, 38, 45, 46]. Such information is relevant, as the stability of the sample should be guaranteed during this time period.

Dried blood sample extraction

Compared to sample preparation protocols of dried blood microsamples, developed for application in therapeutic drug monitoring (TDM), extraction protocols for measurement of HbA1c are more straightforward and require less lab time. For most of the published dried blood based methods, a simple one-step extraction of the sample was performed by adding an amount of a buffer [14, 26], a commercial hemolysing reagent [13, 16, 18, 28], [29], [30], [31, 34], a hemolysis/wash solution [11, 37] or deionized water [9, 32, 38] to the sample, with a subsequent incubation and/or shaking step. The reported elution times ranged from 30 s [9] to 3 h [16]. The elution conditions (volume, time and temperature) were not always reported. Jeppsson et al. used a more elaborate two-step elution protocol by adding a sodium phosphate-citrate buffer containing Triton X-100 as detergent to the DBS punch, followed by the addition of an equal volume of a cysteine containing buffer to eliminate the interfering Hb glutathione adduct [24]. One should be aware that since 2021 the use of Triton X-100 detergent in laboratories is restricted in the European Union.

Analytical methods

The same methods used for measurement of HbA1c in conventional liquid blood samples can be applied on extracted dried blood microsamples. Most of the published dried blood based methods used an automated ion-exchange HPLC analyser: Bio-Rad analyser [13], [14], [15], Arkray ADAMS analyser [26] or Tosoh G series [9, 11, 25, 37, 38, 40], in which HbA1c is separated from other hemoglobin fractions and variants based on different ionic interactions with the stationary phase. Automated HPLC analysers allow rapid and reproducible analysis of samples in a clinical lab, comparable with typical immunoassay workflows. In an older paper, a non-automated HPLC ion-exchange was used [24]. DBS extracts have also been analyzed with turbidimetric immunoassays, in which an anti-HbA1c antibody is used, which reacts with HbA1c to form a complex. Whole blood assay protocols were applied on DBS using different commercially available immunoassays: direct turbidimetric inhibition immunoassay (Thermo Fisher Scientific) [31,33], Tina-quant HbA1c assay (Roche Diagnostics) [16,18,27,28,34], immunoturbidimetric latex method (Agappe Diagnostics) [29], Kamiya Biomedical HbA1c kit [32] and the Olympus HbA1c kit [30]. Two (older) studies reported on the measurement of HbA1c in DBS using affinity chromatography [47,48] and, to the best of our knowledge, none with capillary electrophoresis.

Two mass spectrometry-based HbA1c methods in dried blood matrices were published. In a proof-of-concept study, HbA1c was quantified (and Hb variants identified) in VAMS samples using a data-independent acquisition (DIA) liquid chromatography-mass spectrometry (LC-MS) approach, a technique commonly used in proteomics [39]. A similar method for HbA1c quantification in DBS, together with apoB and apoA-I, using nanoflow LC-parallel reaction monitoring (PRM)-MS was also described [35].

In a few papers, the precision of measuring HbA1c in dried blood microsamples was evaluated. Overall, the obtained coefficients of variation (CVs) were reasonable [16, 26, 32], even though the imprecision of dried blood based methods compared to conventional liquid blood is expected to be higher due to an increased variation in pre-analytical conditions, sample volume and elution efficiency [14]. Hall et al. reported comparable CVs for DBS and fresh blood samples, suggesting a negligible error introduced by DBS collection and elution of the spot [11]. A CV of 5.80 % was obtained for HbA1c duplicate measurements performed on dried VAMS [9] and 2.84 % on wet VAMS samples [38], indicating reproducibility of both sampling techniques (based on results expressed in IFCC units).

Method validation

During method development of a dried blood based analytical method the effect of deviating hematocrits, volume of blood spotted on the DBS card and spot homogeneity (for DBS) on obtained quantitative results should be evaluated [49]. However, since HbA1c values are expressed as a relative ratio (mmol/mol or percentage), these parameters are unlikely to affect the measured HbA1c result. Furthermore, as the standard matrix for HbA1c analysis is whole blood, HbA1c results obtained from dried blood microsamples are directly comparable to liquid blood results. Hence, when compared to for example the application of dried blood microsampling for TDM [10], the validation of HbA1c measurements in dried blood microsamples should be more straightforward.

As dried blood microsamples are typically collected following a finger-prick, capillary blood is collected instead of venous blood. Several studies have demonstrated a good agreement between venous and capillary blood HbA1c results, indicating that capillary and venous blood can be used interchangeably [31, 45, 46, 50], [51], [52], [53]. However, one should be aware that slight differences might be present in patients at the intensive care unit due to oxidative stress [54].

Correlation studies with conventional samples

Most of the studies covered by this review included a method comparison between HbA1c results in the alternative dried blood microsample vs. conventional liquid blood. However, there was a marked heterogeneity in study design, and too often critical evaluation of the data was missing and/or the conclusions were not supported by the presented data. It should also be noted that in a substantial number of these studies DBS were actually not collected via capillary finger-prick, but were artificially laboratory-generated samples, obtained by spotting venous whole blood onto filter paper (see part on sampling conditions above) [13, 15, 16, 25, 26, 29, 30]. Furthermore, in some study designs DBS and whole blood samples were measured in the same run quite shortly after sample collection, which does not mimic a real-life home-sampling situation. Table 1 provides details of the performed correlation studies.

To evaluate the agreement -rather than the correlation-between HbA1c results obtained from dried blood microsamples and venous liquid blood, objective acceptance criteria based on clinical perspectives should be taken into account [55]. The applied acceptance criteria for HbA1c in (dried) blood microsampling methods are summarized in Table 3. For example, in the studies from Verougstraete [9, 38] and Hall [37], Royal College of Pathologists of Australasia (RCPA) quality requirements for total allowable error were used. Alternatively, a maximum acceptable difference of 5 mmol/mol was suggested by Groenendijk et al., as this is the smallest difference in consecutive HbA1c results that guides clinicians to change therapy [56].

Table 3:

Applied acceptance criteria for evaluating the agreement of HbA1c results obtained in (dried) blood microsamples vs. conventional samples.

Acceptance criterion Criterion based on Applied in
Total allowable error: ±4 mmol/mol (IFCC) or ±0.4 % (NGSP) RCPA based on biological variation [63] Verougstraete et al. [9, 38] + Hall et al. [37]
Total allowable relative error of ±6 % in at least 92.5 % of all samples Little RR [64] Elliott et al. [18]
±6 % evaluation limit CAP evaluation limits for proficiency testing Miller et al. [32]
Total allowable relative error of ±5 % in ≥90 % of all samples NGSP [65] Beck et al. [45]
Maximum acceptable difference of ±5 mmol/mol (IFCC) or ±0.5 % (NGSP). Smallest difference in HbA1c values of consecutive tests that guide to change in therapy Groenendijk et al. [56]
95 % confidence interval of the differences must be within ±0.75 % (NGSP) or ±8.2 mmol/mol (IFCC) HbA1c NGSP [66] Jones et al. [16]
Total allowable error: ±5 mmol/mol (IFCC) or ±0.5 % (NGSP) Weykamp et al. [67] Verougstraete et al. [9] + clinical laboratory of UZ Ghent
Total allowable relative error of ±4.6 % Biological variation of HbA1c [68] Crawford et al. [34]
  1. If absolute HbA1c values were used, the values are shown in both IFCC (mmol/mol) and NGSP (%) units. CAP, College of American Pathologists; IFCC, International Federation of Clinical Chemistry; NGSP, National Glycohemoglobin Standardization Program; RCPA, Royal College of Pathologists of Australasia.

Most studies reported a high correlation between HbA1c results obtained from DBS and venous liquid whole blood [13, 14, 16, 24, 28], [29], [30, 32]. However, a critical evaluation of the obtained data, using one of the formal acceptance criteria in Table 3, was lacking in several instances. Several studies applied a conversion formula to adjust DBS results to yield a better correlation (or agreement) with traditional venous blood levels [11, 15, 18, 26, 31, 33]. However, when working with such experimentally determined conversion factor or formula, it is important that the utility of such a conversion is evaluated on an independent cohort, and not only on the cohort which was used for the set-up of the factor or formula [49]. This can also be achieved by randomizing samples into two distinct pools: one to set up a calibration model, and the other for evaluation of the model or correction factor [55]. For example, Scherer et al. measured 14,243 DBS cards to set up a regression analysis to correct for storage time, and validated the predictive performance of this model using 4,272 distinct samples from the same participants [33]. Crawford et al. set up an initial comparison study with assisted finger-prick sampling to establish a calibration factor to convert finger-prick HbA1c results to venous values using a “training dataset” of more than 100 samples. Subsequently, the obtained assay-specific calibration factor was independently verified on a smaller “testing dataset” including participants’ self-collected microsamples [34].

The time elapsed between sample collection and analysis does seem to be a key factor impacting HbA1c DBS results. Mastronardi et al. observed that the correlation between DBS and venous blood HbA1c values became weaker over time: whereas results were still acceptable up to 7 days after collection (storage conditions were not mentioned), the obtained differences were clinically unacceptable after 14 days. Moreover, the (analytical) variability increased in older samples. Multiple correction equations were set-up, for each particular day of sample analysis after collection, and applied to predict HbA1c whole blood values from DBS measurements [31]. Similarly, Elliott et al. observed that the obtained biases on HbA1c measured in DBS samples increased with time between collection and analysis. DBS samples received more than 7 days after collection were excluded because of the expectation of significant and unpredictable errors [18]. Hall et al. evaluated the impact of time between HemaSpot DBS preparation and HbA1c analysis. DBS HbA1c results, collected with HemaSpot™ devices, analyzed up to five days after collection (stored at ambient temperature) were compared with capillary liquid blood HbA1c results. After applying a day-dependent correction formula, home-prepared DBS HbA1c results were stated to be equivalent to the liquid blood results [37]. Remarkably, in their previous study the authors stated that the time between HemaSpot™ DBS preparation and analysis (up to 4 days) did not affect the HbA1c results and the regression model. However, the validity of that statement can be questioned as over 80 % of the samples were analyzed within 2 days and only 2 (out of 104) samples were analyzed 4 days after collection [11]. Obviously, application of a correction equation yielded better results in all the above-mentioned studies. However, an important limitation is that in none of these studies the equations were verified on independent datasets. This is also illustrated in the Hall study [37], where the predicted HbA1c values, based on the day-dependent regression equations, correlated better than the ones obtained using a more generic, previously published formula [11]. This stresses once more the need for critical data evaluation.

In a recent multicenter study the impact of different DBS storage conditions and analysis delay on HbA1c measurement validity was evaluated. Delay in analysis ranged from 0 to 400 days; with 50.4 % of the samples being analyzed within the first week after collection and approximately 90 % within 4 weeks. Overall, HbA1c levels increased with increasing analysis delay. A correction method, taking into account the delay (in days) and a center-specific variable, reflecting different storage conditions per participating center, was set up. Application of this method on an independent set of DBS led to a significant improvement of the correlation of the DBS HbA1c results with the corresponding whole blood values [33].

In a recent study, Verougstraete et al. compared HbA1c results in conventional liquid blood samples to results obtained in DBS and VAMS samples collected at home by the patients. Poor agreements were found between DBS/VAMS HbA1c results and the liquid blood results and too many samples did not meet the RCPA quality requirements. Interestingly, higher agreements with liquid blood samples were obtained for DBS and VAMS samples that were analyzed 1–3 days after collection compared to those that were analyzed 4–6 days after collection. A possible explanation for this was that by ageing of the dried samples the hemoglobin components were converted to oxidized and degraded hemoglobin which had an unpredictable effect on the HbA1c result, making it impossible to set-up a bias correction equation [9]. In another study it was also observed that aged liquid whole blood samples often yielded unacceptable chromatograms by the appearance of unidentified extra peaks, possibly corresponding to hemoglobin degradation products [42]. Another study, analyzing HbA1c with LC-MS in samples collected by VAMS, found that upon storage at room temperature, HbA1c values increased. This was a specific limitation of glycated peptides, since the non-glycated evaluated peptides showed stability up to 3 or 10 days at room temperature [39]. Besides the overall good performance obtained during validation of another LC-MS method on DBS, the developed method was considered not clinically useful as the authors observed sample-specific disagreement compared to the conventional validated HbA1c assay on venous samples [35]. Degradation and conversion of hemoglobin products during drying, transport and storage of dried samples has also been described in other reports. In particular ion-exchange chromatography-based methods may struggle with these conversions, which yield abnormal elution patterns, as described above. Moreover, degradation products have the potential to co-elute or may be incompletely separated from the HbA1c peak, both potentially leading to erroneous HbA1c results [9, 13, 26]. This was nicely exemplified by a study conducted by Thomas et al. These authors obtained DBS HbA1c values that were substantially and significantly higher compared to the corresponding liquid blood values using ion-exchange HPLC. This upward bias was considered clinically relevant, with potential overestimation of diabetes prevalence rates. Interestingly, it was observed by the authors that their HPLC ion-exchange technique failed to capture a part of the total Hb fraction that should be included in the denominator of the %HbA1c calculation. This was due to the presence of an extra unexpected peak to the right of A0, probably misclassified by the software as a Hb variant. The origin of this peak could not be found. After recalculation by including chromatographic areas that were initially ignored, HbA1c values from DBS closely agreed with the corresponding venous values [15]. Whereas the formed hemoglobin degradation products cause interfering peaks on ion-exchange methods because of their altered ionic charges, it is expected that boronate affinity HPLC methods would not be affected, as the cis-diol group on the glycated hemoglobin is not changed [43]. To the best of our knowledge, only two (older) filter paper-based methods were published using this methodology [47, 48] and measuring HbA1c with boronate affinity analysers is not widespread in clinical labs because of the limited availability of automated analysers. Another possibility is using turbidimetric immunoassays for measuring dried blood microsamples as these might be more robust towards the formed hemoglobin conversions, however, also these assays yielded deviating DBS HbA1c values [18, 28, 31, 34].

In their study, Verougstaete et al. also evaluated the effect of analyzing VAMS samples immediately after collection, with the samples being capped with a shell to delay the drying process. The obtained HbA1c results on these VAMS samples showed an excellent agreement with the corresponding liquid capillary blood sample and all samples fulfilled the RCPA requirements. This observation showed the intrinsic potential of using VAMS as a sampling technique for obtaining clinically acceptable HbA1c results [9]. Based on the latter finding, a follow-up study was set-up in diabetic children who used VAMS for home-sampling. However, instead of drying the samples, the VAMS tips were deposited in 800 µL sterile water shortly after collection (within 30 min). The samples were also stored and sent to the lab in water at ambient temperature. Overall, this approach showed a substantial improvement over the dried VAMS approach and the apparent hemoglobin conversion taking place during drying, and impacting the HPLC ion-exchange analysis, was possibly prevented [38].

Patient satisfaction

In several studies the patient satisfaction was evaluated by taking questionnaires. Overall, patients were satisfied with the alternative sampling techniques and indicated that the sample collection was easy to perform and that they preferred home sampling over traditional venous sampling at a medical center [11, 18, 28]. A recurrent concern from subjects was the difficulty of ensuring that there was enough blood on the DBS paper. To obtain high quality samples collected by the patients themselves, proper instruction (preferably with written instructions containing pictures) prior to collection is essential. In a study which combined DBS and VAMS sampling, participants experienced both sampling techniques as very convenient, with VAMS standing out as the preferred sampling technique (especially in the pediatric population) [9, 38].

Liquid blood microsampling

As an alternative to dried blood microsampling, also liquid whole blood microsampling seems to be a promising option. The same advantages as with dried blood microsamples are to be expected: samples can be collected by patients via a minimally invasive finger-prick, after which they can be sent to the lab via regular post. Shipping and storage is still considered more convenient compared to conventional sampling because of the low sample volumes. However, one should be aware that liquid samples have an intrinsic higher biohazard risk and inherently carry the potential of leaking during transport, requiring suitable (leakproof) containers. Liquid capillary blood specimens were already remotely collected and sent to the lab for measurement of HbA1c used as assistance for virtual consultations during the COVID pandemic [45, 56, 57]. Capillary blood was collected into suitable calibrated/precision (heparinized) capillaries [45, 46, 50] or small EDTA anticoagulated MiniCollect™ [56], Microtainer™ [57] or Hem-Col™ [58] tubes. After collection, the capillary tubes were stored in a stabilizing or hemolyzing solution and transported as such to the lab [45, 50, 59]. Comparable to dried samples, also for the measurement of HbA1c in liquid microsamples different analytical techniques were used: ion-exchange HPLC [45, 46, 50, 58], capillary electrophoresis [56] and a point of care (POC) immunoassay-based analyser [57]. Compared to dried blood microsamples, no stability issues have been reported, suggesting that this may be a suitable approach. Table 4 gives an overview of the published methods using liquid blood microsamples.

Table 4:

Overview of published liquid blood microsample HbA1c-based methods.

Reference Analytical technique Application + included patients/samples Sample collection Time between collection and analysis Specific remarks
Beck et al. [45] Ion-exchange HPLC (Tosoh G8) Comparison between liquid capillary and venous blood from 240 diabetes patients 5–15 µL capillary blood collected by the participants (or their parents) in the clinic unsupervised Up to 10 days Extensive study on the use of liquid blood microsamples as alternative for venous specimens
Huijskens et al. [58] Ion-exchange HPLC (Tosoh G8) Comparison between liquid capillary and venous blood from random patients requiring blood collection 200 µL capillary (≈4–5 drops) blood collected by experienced phlebotomists Up to 5 days
Nathan et al. [46] Ion-exchange HPLC (Tosoh) Comparison between liquid capillary and venous blood from 122 diabetes patients Capillary blood collected by the participants in the clinic Up to 22 days
Voss et al. [50] Ion-exchange HPLC (Bio-Rad Diamat) Comparison between venous blood and liquid capillary blood from 47 pediatric diabetes patients 5 µL capillary blood at the hospital Up to 22 days
Groenendijk et al. [56] Capillary electrophoresis (Sebia Capillarys 3 Tera) Comparison between liquid capillary and venous blood from 41 diabetes patients 20 µL capillary blood collected by the participants remotely ?
Cross et al. [57] POC analyser (Siemens DCA Vantage) Comparison between 286 capillary blood samples collected at the diabetes center and at home by diabetes patients 50–100 µL capillary blood (≈5–10 drops) collected at the participant’s home Up to 12 days Evaluation of at-home blood sampling method for measuring HbA1c for virtual consultations
  1. Papers are ordered per analytical technique and in alphabetical order of the first author. Papers of special interest/relevance have been highlighted in bold.

An excellent agreement between capillary blood and venous blood specimens from 47 pediatric diabetes patients was obtained. After collection, calibrated capillary tubes were placed into vials containing a stabilizing solution of EDTA and KCl [50]. The same researchers also demonstrated the feasibility of mailing such capillary blood samples to the lab at ambient temperatures [60]. Similar results were reported by Cross et al.: diabetes patients collected 50 µL capillary blood (5 drops) in Microtainer™ tubes at home and sent these to the lab via postal services. Clinically acceptable results compared to capillary samples taken at the diabetes center were obtained [57]. In another study, almost identical HbA1c values were observed in 41 paired capillary (collected at home by the participants in MiniCollect™ tubes and subsequently mailed to the lab) and venous (collected during clinic appointment) samples: 40/41 of the capillary results were within the total allowable error of ±5 mmol/mol [56].

Commercial kits for collection of these capillary samples at home already exist. For example the Bio-Rad Hemoglobin Capillary Collection System [61] consists of a 5 µL plastic capillary which should be filled with capillary blood from a fingertip. Thereafter, the vial should be transferred to a sample preparation vial, containing an aqueous solution of EDTA and potassium cyanide (0.25 mmol/L). After capping, the vial should be shaken to rinse the blood completely from the capillary into the solution. After that, the sample is ready for analysis, transport or storage. According to the insert, samples are stable for two weeks at room temperature, four weeks at 2–8 °C and four days at 42 °C [61]. These collection kits were used in two recent studies and were found suitable for HbA1c monitoring at home with shipping of the samples to a central laboratory. In the first study, liquid capillary blood samples were compared with corresponding venous blood samples obtained from 240 diabetes patients (adult and pediatric). The Bio-Rad kit was compared to another kit containing a microfuge tube filled with a stabilizing solution of EDTA and other undisclosed preservatives. Capillary results from both collection kits agreed with the conventional results as more than 95 % of the results were within ±5 % of the venous HbA1c value. Moreover, no differences in HbA1c results were observed between samples shipped with or without a cold pack by USPS, or shipped on dry ice by overnight carrier. HbA1c results were also not affected by temperatures during the shipment exceeding 37.8 °C and by shipping times up to 10 days [45]. Nathan and colleagues confirmed the ease and possibility to obtain clinically acceptable HbA1c results from capillary finger-prick samples collected using the Bio-Rad collection kit [46].

Liquid capillary microsamples were found to be stable for up to 7 days when stored at 4 °C [56] and up to 12 days when stored at room temperature [57], allowing postal delays. Using Hem-Col™ devices (microtubes containing EDTA dissolved in preservation fluid), HbA1c was found to be stable for at least 5 days at room temperature, 4 °C and 37 °C [58]. Liquid capillary blood remote collections were also positively evaluated by the users and were perceived as easy to use [45, 56]. At our hospital, collecting capillary blood in a heparinized capillary tube is currently still the standard procedure for measuring HbA1c in pediatric population [9].

In a recent published systematic review it was concluded that for self-collected and posted capillary blood microsamples, HbA1c results in microtubes/capillary tubes agreed better with conventional venous values than dried blood microsamples. Moreover, only the liquid microsample based methods met IFCC quality targets (i.e. ±5 mmol/mol), and the limits of agreement, when compared to venous liquid blood, were smaller than those observed when using dried blood microsamples [62].

Conclusions

Over the past four decades, a plethora of papers has been published on the measurement of HbA1c in blood microsamples. However, the published data on study designs, results and conclusions are considerably heterogeneous. Moreover, many of the included studies lacked details on collection, storage and shipment conditions of the dried samples. As a consequence, different conclusions from apparently similar results were drawn. To promote implementation of liquid or dried blood microsampling based approaches in routine clinical care – which is to our knowledge still sparse – the entire procedure (from collection to analysis) should be ‘under control’. To achieve successful implementation in the clinical laboratory, a few issues should be encountered during method validation. First, it is important that the expected conditions of the sample collection of the future intended method are evaluated. If for example the intended goal is home-sampling, microsamples actually collected by the participants in their home environment must be compared to the conventional blood samples. Expected conditions encountered during (postal) transportation of the microsamples to the lab should also be mimicked in the validation experiments. The study population should also resemble real-life patients. Moreover, when evaluating the agreement between HbA1c results obtained from microsamples and traditional venous liquid blood, objective acceptance criteria should be used. Lastly, if a conversion factor or equation is used to correct the obtained microsample results, this should be evaluated on an independent cohort and not on the dataset which was used to generate the factor or equation.

Despite the heterogeneity of the findings from the studies included in this review, it is clear that there is an issue with drying and ageing of the microsamples to obtain clinically acceptable HbA1c results. Instead of using dried blood microsamples, liquid microsamples may be a better alternative to remotely monitor HbA1c: high agreements with HbA1c results from conventional samples have been described and no stability issues have been reported. Several commercial sampling kits are already available to collect and transport capillary liquid blood in a convenient manner.


Corresponding author: Professor Christophe P. Stove, Laboratory of Toxicology, Department of Bioanalysis, Faculty of Pharmaceutical Sciences, Ghent University; Ottergemsesteenweg 460, 9000 Gent, Belgium, E-mail:

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Not applicable.

References

1. American Diabetes Association Professional Practice Committee. Classification and diagnosis of diabetes: standards of medical care in diabetes-2022. Diabetes Care 2022;45:S17–38. https://doi.org/10.2337/dc22-s002.Search in Google Scholar PubMed

2. The International Expert Committee. International Expert Committee report on the role of the A1C assay in the diagnosis of diabetes. Diabetes Care 2009;32:1327–34. https://doi.org/10.2337/dc09-9033.Search in Google Scholar PubMed PubMed Central

3. Zhang, X, Gregg, EW, Williamson, DF, Barker, LE, Thomas, W, Bullard, KM, et al.. A1C level and future risk of diabetes: a systematic review. Diabetes Care 2010;33:1665–73. https://doi.org/10.2337/dc09-1939.Search in Google Scholar PubMed PubMed Central

4. American Diabetes Association Professional Practice Committee. Glycemic targets: standards of medical care in diabetes—2022. Diabetes Care 2021;45:S83–S96. https://doi.org/10.2337/dc22-s006.Search in Google Scholar PubMed

5. Gillery, P. HbA(1c) and biomarkers of diabetes mellitus in clinical chemistry and laboratory medicine: ten years after. Clin Chem Lab Med 2022;61:861–72. https://doi.org/10.1515/cclm-2022-0894.Search in Google Scholar PubMed

6. National Glycohemoglobin Standardization Program. Certified methods and laboratories. Available from: http://ngsp.org/certified.asp [Accessed Sept 2022].Search in Google Scholar

7. Little, RR, Roberts, WL. A review of variant hemoglobins interfering with hemoglobin A1c measurement. J Diabetes Sci Technol 2009;3:446–51. https://doi.org/10.1177/193229680900300307.Search in Google Scholar PubMed PubMed Central

8. National Glycohemoglobin Standardization Program. HbA1c assay interferences. Available from: http://www.ngsp.org/interf.asp [Accessed Sept 2022].Search in Google Scholar

9. Verougstraete, N, Lapauw, B, Van Aken, S, Delanghe, J, Stove, C, Stove, V. Volumetric absorptive microsampling at home as an alternative tool for the monitoring of HbA1c in diabetes patients. Clin Chem Lab Med 2017;55:462–9. https://doi.org/10.1515/cclm-2016-0411.Search in Google Scholar PubMed

10. Verougstraete, N, Stove, V, Verstraete, AG, Stove, CP. Therapeutic drug monitoring of tyrosine kinase inhibitors using dried blood microsamples. Front Oncol 2022;12:821807. https://doi.org/10.3389/fonc.2022.821807.Search in Google Scholar PubMed PubMed Central

11. Hall, JM, Fowler, CF, Barrett, F, Humphry, RW, Van Drimmelen, M, MacRury, SM. HbA(1c) determination from HemaSpot™ blood collection devices: comparison of home prepared dried blood spots with standard venous blood analysis. Diabet Med 2020;37:1463–70. https://doi.org/10.1111/dme.14110.Search in Google Scholar PubMed PubMed Central

12. Schnell, O, Crocker, JB, Weng, J. Impact of HbA1c testing at point of care on diabetes management. J Diabetes Sci Technol 2017;11:611–7. https://doi.org/10.1177/1932296816678263.Search in Google Scholar PubMed PubMed Central

13. Egier, DA, Keys, JL, Hall, SK, McQueen, MJ. Measurement of hemoglobin A1c from filter papers for population-based studies. Clin Chem 2011;57:577–85. https://doi.org/10.1373/clinchem.2010.156380.Search in Google Scholar PubMed

14. Lacher, DA, Berman, LE, Chen, TC, Porter, KS. Comparison of dried blood spot to venous methods for hemoglobin A1c, glucose, total cholesterol, high-density lipoprotein cholesterol, and C-reactive protein. Clin Chim Acta 2013;422:54–8. https://doi.org/10.1016/j.cca.2013.03.032.Search in Google Scholar PubMed

15. Thomas, D, Seeman, T, Potter, A, Hu, P, Crimmins, E, Herningtyas, EH, et al.. HPLC-Based measurement of glycated hemoglobin using dried blood spots collected under adverse field conditions. Biodemogr Soc Biol 2018;64:43–62. https://doi.org/10.1080/19485565.2018.1451300.Search in Google Scholar PubMed PubMed Central

16. Jones, TG, Warber, KD, Roberts, BD. Analysis of hemoglobin A1c from dried blood spot samples with the Tina-quantR II immunoturbidimetric method. J Diabetes Sci Technol 2010;4:244–9. https://doi.org/10.1177/193229681000400203.Search in Google Scholar PubMed PubMed Central

17. Wikblad, K, Smide, B, Bergström, A, Wahren, L, Mugusi, F, Jeppsson, JO. Immediate assessment of HbA1c under field conditions in Tanzania. Diabetes Res Clin Pract 1998;40:123–8. https://doi.org/10.1016/s0168-8227(98)00046-1.Search in Google Scholar PubMed

18. Elliott, TG, Dooley, KC, Zhang, M, Campbell, HSD, Thompson, DJS. Comparison of glycated hemoglobin results based on at-home and in-lab dried blood spot sampling to routine venous blood sampling in-lab in adult patients with type 1 or type 2 diabetes. Can J Diabetes 2018;42:426–32.e1. https://doi.org/10.1016/j.jcjd.2017.10.053.Search in Google Scholar PubMed

19. Protti, M, Mandrioli, R, Mercolini, L. Quantitative microsampling for bioanalytical applications related to the SARS-CoV-2 pandemic: usefulness, benefits and pitfalls. J Pharmaceut Biomed Anal 2020;191:113597. https://doi.org/10.1016/j.jpba.2020.113597.Search in Google Scholar PubMed PubMed Central

20. Wong, MP, Meas, MA, Adams, C, Hernandez, S, Green, V, Montoya, M, et al.. Development and implementation of dried blood spot-based COVID-19 serological assays for epidemiologic studies. Microbiol Spectr 2022;10:e0247121. https://doi.org/10.1128/spectrum.02471-21.Search in Google Scholar PubMed PubMed Central

21. Roberts, AJ, Malik, F, Pihoker, C, Dickerson, JA. Adapting to telemedicine in the COVID-19 era: feasibility of dried blood spot testing for hemoglobin A1c. Diabetes Metabol Syndr 2021;15:433–7. https://doi.org/10.1016/j.dsx.2021.02.010.Search in Google Scholar PubMed PubMed Central

22. Garnett, ER, Recio, B, Jung, J, Tam, E, Devaraj, S. Pandemic-associated trends in measurement of HbA1c in children with diabetes mellitus and validation of dried blood spot as an alternative sample matrix. Ann Clin Lab Sci 2021;51:535–9.Search in Google Scholar

23. Gill, EL, Patel, K, Dickerson, JA, Dulik, MC, Grant, RP, Heaney, DL, et al.. Alternative sample matrices supporting remote sample collection during the pandemic and beyond. Clin Chem 2022;68:269–75. https://doi.org/10.1093/clinchem/hvab257.Search in Google Scholar PubMed

24. Jeppsson, JO, Jerntorp, P, Almër, LO, Persson, R, Ekberg, G, Sundkvist, G. Capillary blood on filter paper for determination of HbA1c by ion exchange chromatography. Diabetes Care 1996;19:142–5. https://doi.org/10.2337/diacare.19.2.142.Search in Google Scholar PubMed

25. Pollock, AJ, Allen, DB, Wiebe, D, Eickhoff, J, MacDonald, M, Baker, M. Development of filter paper hemoglobin A1c assay applicable to newborn screening. Clin Chim Acta 2016;457:24–6. https://doi.org/10.1016/j.cca.2016.03.014.Search in Google Scholar PubMed PubMed Central

26. Maleska, A, Hirtz, C, Casteleyn, E, Villard, O, Ducos, J, Avignon, A, et al.. Comparison of HbA1c detection in whole blood and dried blood spots using an automated ion-exchange HPLC system. Bioanalysis 2017;9:427–34. https://doi.org/10.4155/bio-2016-0278.Search in Google Scholar PubMed

27. Parkes, J, Ray, R, Kerestan, S, Davis, H, Ginsberg, B. Prospective evaluation of accuracy, precision, and reproducibility of an at-home hemoglobin A1c sampling kit. Diabetes Technol Therapeut 1999;1:411–9. https://doi.org/10.1089/152091599316928.Search in Google Scholar PubMed

28. Fokkema, MR, Bakker, AJ, de Boer, F, Kooistra, J, de Vries, S, Wolthuis, A. HbA1c measurements from dried blood spots: validation and patient satisfaction. Clin Chem Lab Med 2009;47:1259–64. https://doi.org/10.1515/cclm.2009.274.Search in Google Scholar

29. Lakshmy, R, Gupta, R. Measurement of glycated hemoglobin A1c from dried blood by turbidimetric immunoassay. J Diabetes Sci Technol 2009;3:1203–6. https://doi.org/10.1177/193229680900300527.Search in Google Scholar PubMed PubMed Central

30. Hu, P, Edenfield, M, Potter, A, Kale, V, Risbud, A, Williams, S, et al.. Validation and modification of dried blood spot-based glycosylated hemoglobin assay for the longitudinal aging study in India. Am J Hum Biol 2015;27:579–81. https://doi.org/10.1002/ajhb.22664.Search in Google Scholar PubMed PubMed Central

31. Mastronardi, CA, Whittle, B, Tunningley, R, Neeman, T, Paz-Filho, G. The use of dried blood spot sampling for the measurement of HbA1c: a cross-sectional study. BMC Clin Pathol 2015;15:13. https://doi.org/10.1186/s12907-015-0013-5.Search in Google Scholar PubMed PubMed Central

32. Miller, IM, Lacher, DA, Chen, TC, Zipf, GW, Gindi, RM, Galinsky, AM, et al.. Collection and laboratory methods for dried blood spots for hemoglobin A1c and total and high-density lipoprotein cholesterol in population-based surveys. Clin Chim Acta 2015;445:143–54. https://doi.org/10.1016/j.cca.2015.03.028.Search in Google Scholar PubMed PubMed Central

33. Scherer, N, Kurbasic, A, Dings, C, Mari, A, Nock, V, Hennige, AM, et al.. The impact and correction of analysis delay and variability in storage temperature on the assessment of HbA1c from dried blood spots – an IMI direct study. Int J Proteom Bioinf 2019;4:7–13.Search in Google Scholar

34. Crawford, ML, Collier, BB, Bradley, MN, Holland, PL, Shuford, CM, Grant, RP. Empiricism in microsampling: utilizing a novel lateral flow device and intrinsic normalization to provide accurate and precise clinical analysis from a finger stick. Clin Chem 2020;66:821–31. https://doi.org/10.1093/clinchem/hvaa082.Search in Google Scholar PubMed

35. Henderson, CM, Bollinger, JG, Becker, JO, Wallace, JM, Laha, TJ, MacCoss, MJ, et al.. Quantification by nano liquid chromatography parallel reaction monitoring mass spectrometry of human apolipoprotein A-I, apolipoprotein B, and hemoglobin A1c in dried blood spots. Proteomics. Clin Appl 2017;11. https://doi.org/10.1002/prca.201600103.Search in Google Scholar PubMed PubMed Central

36. Goldstein, DE, Wiedmeyer, HM, England, JD, Little, RR, Parker, KM. Glycosylated protein in whole blood spotted on filter paper. Clin Chem 1982;28:386–7. https://doi.org/10.1093/clinchem/28.2.386.Search in Google Scholar

37. Hall, JM, Fowler, CF, Pollock, MA, MacRury, SM. Haemoglobin A1c determination from dried blood spots prepared with HemaSpot™ blood collection devices: comparison with fresh capillary blood. Clin Chem Lab Med 2020;59:e79–82. https://doi.org/10.1515/cclm-2020-0675.Search in Google Scholar PubMed

38. Verougstraete, N, Stove, V, Stove, C. Wet absorptive microsampling at home for HbA1c monitoring in diabetic children. Clin Chem Lab Med 2018;56:e291–4. https://doi.org/10.1515/cclm-2018-0207.Search in Google Scholar PubMed

39. Lima, DA, Schuch, RA, Salgueiro, JS, Pintão, MCT, Carvalho, VM. Evaluation of volumetric absorptive microsampling and mass spectrometry data-independent acquisition of hemoglobin-related clinical markers. J Proteome Res 2022;21:1816–28. https://doi.org/10.1021/acs.jproteome.1c00892.Search in Google Scholar PubMed

40. Acar, K, Kurtulus Dereli, A, Avci, E, Zeybek, V, Kutlu, E, Demir, S, et al.. Determination of haemoglobin A1c levels using high-performance liquid chromatography of bloodstains. Med Sci Law 2020;60:19–25. https://doi.org/10.1177/0025802419879272.Search in Google Scholar PubMed

41. Ghent University. VID 20190115 153239. Available from: https://www.youtube.com/watch?v=5nCyNUFCF3c [Accessed Dec 2022].Search in Google Scholar

42. Little, RR, Rohlfing, CL, Tennill, AL, Connolly, S, Hanson, S. Effects of sample storage conditions on glycated hemoglobin measurement: evaluation of five different high performance liquid chromatography methods. Diabetes Technol Therapeut 2007;9:36–42. https://doi.org/10.1089/dia.2006.0055.Search in Google Scholar PubMed

43. Rohlfing, CL, Hanson, S, Tennill, AL, Little, RR. Effects of whole blood storage on hemoglobin a1c measurements with five current assay methods. Diabetes Technol Therapeut 2012;14:271–5. https://doi.org/10.1089/dia.2011.0136.Search in Google Scholar PubMed PubMed Central

44. Affan, ET, Praveen, D, Chow, CK, Neal, BC. Comparability of HbA1c and lipids measured with dried blood spot versus venous samples: a systematic review and meta-analysis. BMC Clin Pathol 2014;14:21. https://doi.org/10.1186/1472-6890-14-21.Search in Google Scholar PubMed PubMed Central

45. Beck, RW, Bocchino, LE, Lum, JW, Kollman, C, Barnes-Lomen, V, Sulik, M, et al.. An evaluation of two capillary sample collection kits for laboratory measurement of HbA1c. Diabetes Technol Therapeut 2021;23:537–45. https://doi.org/10.1089/dia.2021.0023.Search in Google Scholar PubMed

46. Nathan, DM, Krause-Steinrauf, H, Braffett, BH, Arends, VL, Younes, N, McGee, P, et al.. Comparison of central laboratory HbA1c measurements obtained from a capillary collection versus a standard venous whole blood collection in the GRADE and EDIC studies. PLoS One 2021;16:e0257154. https://doi.org/10.1371/journal.pone.0257154.Search in Google Scholar PubMed PubMed Central

47. Little, RR, McKenzie, EM, Wiedmeyer, HM, England, JD, Goldstein, DE. Collection of blood on filter paper for measurement of glycated hemoglobin by affinity chromatography. Clin Chem 1986;32:869–71. https://doi.org/10.1093/clinchem/32.5.869.Search in Google Scholar

48. Gay, EC, Cruickshanks, KJ, Chase, HP, Klingensmith, G, Hamman, RF. Accuracy of a filter paper method for measuring glycosylated hemoglobin. Diabetes Care 1992;15:108–10. https://doi.org/10.2337/diacare.15.1.108.Search in Google Scholar PubMed

49. Capiau, S, Veenhof, H, Koster, RA, Bergqvist, Y, Boettcher, M, Halmingh, O, et al.. Official international association for therapeutic drug monitoring and clinical toxicology guideline: development and validation of dried blood spot-based methods for therapeutic drug monitoring. Ther Drug Monit 2019;41:409–30. https://doi.org/10.1097/ftd.0000000000000643.Search in Google Scholar PubMed

50. Voss, EM, Cembrowski, GS, Clasen, BL, Spencer, ML, Ainslie, MB, Haig, B. Evaluation of capillary collection system for HbA1c specimens. Diabetes Care 1992;15:700–1. https://doi.org/10.2337/diacare.15.5.700.Search in Google Scholar PubMed

51. Shah, AR, Challener, J, Elsey, TS, Maguire, GA, Calvin, J, Rayman, G. A novel capillary collection method for obtaining current glycosylated haemoglobin levels in diabetic children. Diabet Med 1994;11:319–22. https://doi.org/10.1111/j.1464-5491.1994.tb00278.x.Search in Google Scholar PubMed

52. Lytken Larsen, M. Evaluation of a new capillary blood collection system for laboratory assay of glycated haemoglobin. Scand J Clin Lab Investig 1986;46:315–7. https://doi.org/10.3109/00365518609083676.Search in Google Scholar PubMed

53. Heylen, O, Van Neyghem, S, Exterbille, S, Wehlou, C, Gorus, F, Weets, I. Evaluation of the Sebia CAPILLARYS 2 flex piercing for the measurement of HbA(1c) on venous and capillary blood samples. Am J Clin Pathol 2014;141:867–77. https://doi.org/10.1309/ajcpru5qc2jbansv.Search in Google Scholar

54. Godefroid, MJ, De Buyzere, ML, Delanghe, JR. Interchangeability of venous and capillary HbA(1c) results is affected by oxidative stress. Clin Chem Lab Med 2013;51:e9–11. https://doi.org/10.1515/cclm-2012-0312.Search in Google Scholar PubMed

55. Verougstraete, N, Stove, V, Stove, C. Comment on HbA(1c) determination from HemaSpot blood collection devices: comparison of home-prepared dried blood spots with standard venous blood analysis. Diabet Med 2020;37:1613–4. https://doi.org/10.1111/dme.14200.Search in Google Scholar PubMed

56. Groenendijk, WN, Griffin, TP, Islam, MN, Blake, L, Wall, D, Bell, M, et al.. Remote capillary blood collection for HbA(1c) measurement during the COVID-19 pandemic: a laboratory and patient perspective. Diabet Med 2022;39:e14897. https://doi.org/10.1111/dme.14897.Search in Google Scholar PubMed PubMed Central

57. Cross, J, Sharma, S, John, WG, Rayman, G. Validation and feasibility of a postal system for remote monitoring of HbA1c. BMJ Open Diabetes Res Care 2021;9. https://doi.org/10.1136/bmjdrc-2021-002527.Search in Google Scholar PubMed PubMed Central

58. Huijskens, M, Castel, R, Vermeer, HJ, Verheijen, FM. Evaluation of Diabetes Care parameters in capillary blood collected with a novel sampling device. Practical Lab Med 2019;17:e00135. https://doi.org/10.1016/j.plabm.2019.e00135.Search in Google Scholar PubMed PubMed Central

59. Bendavid, C, Peltier, L, Letellier, C. Capillary blood sampling kit for HbA1c versus venous puncture on Capillarys 2 flex piercing. Clin Chem Lab Med 2015;53(Suppl):S655.Search in Google Scholar

60. Voss, EM, Cembrowski, GS, Haig, B, Spencer, ML. Stability of mailed and couriered capillary HbA1c samples. Diabetes Care 1993;16:665–6. https://doi.org/10.2337/diacare.16.4.665.Search in Google Scholar PubMed

61. Bio-Rad. Hemoglobin capillary collection system. Available from: https://www.bio-rad.com/en-be/product/hemoglobin-capillary-collection-system?ID=676dfa9b-8395-42b4-9362-0a47c38401dc [Accessed Sept 2022].Search in Google Scholar

62. Colley, J, Dambha-Miller, H, Stuart, B, Bartholomew, J, Benton, M, Baykoca, J, et al.. Home monitoring of Haemoglobin A1c in diabetes: a systematic review and narrative synthesis on accuracy, reliability, and patient acceptability. Diabet Med 2023;40:e15033. e15033. https://doi.org/10.1111/dme.15033.Search in Google Scholar PubMed

63. Royal College of Pathologists of Australasia. RCPA quality requirements. Available from: https://www.westgard.com/rcpa.htm [Accessed Sept 2022].Search in Google Scholar

64. Little, RR. Performance of hemoglobin A1c assay methods: good enough? Clin Chem 2014;60:1031–3. https://doi.org/10.1373/clinchem.2014.225789.Search in Google Scholar PubMed PubMed Central

65. National Glycohemoglobin Standardization Program. Obtaining certification – summary of NGSP criteria. Available from: http://www.ngsp.org/critsumm.asp [Accessed Sept 2022].Search in Google Scholar

66. National Glycohemoglobin Standardization Program. NGSP steering committee meeting; 2010. Available from: http://www.ngsp.org/SC2010.asp [Accessed Sept 2022].Search in Google Scholar

67. Weykamp, CW, Mosca, A, Gillery, P, Panteghini, M. The analytical goals for hemoglobin A(1c) measurement in IFCC units and National Glycohemoglobin Standardization Program Units are different. Clin Chem 2011;57:1204–6. https://doi.org/10.1373/clinchem.2011.162719.Search in Google Scholar PubMed

68. Westgard, QC. Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation. Available from: https://www.westgard.com/biodatabase1.htm [Accessed Sept 2022].Search in Google Scholar

Received: 2023-03-02
Accepted: 2023-06-06
Published Online: 2023-07-10
Published in Print: 2024-01-26

© 2023 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 9.5.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2023-0228/html