Abstract
Objectives
To assess the usefulness of the EP31-A-IR guideline published by the Clinical and Laboratory Standards Institute (CLSI) to perform the periodic verification of results’ comparability between several analyzers.
Methods
Twenty-four biochemistry parameters that could be measured in different analyzers were included: albumin, alkaline phosphatase, alanine aminotransferase, amylase, aspartate aminotransferase, calcium, chloride, C-reactive protein, creatine kinase, creatinine, direct bilirubin, gamma glutamyl transferase, glucose, lactate dehydrogenase, magnesium, phosphate, potassium, sodium, total bilirubin, total cholesterol, total protein, triglycerides, urea and uric acid. In accordance with the EP31-A-IR guideline: (1) Patient samples were selected considering the concentration or activity of interest. (2) Acceptance criteria were established specifically for each concentration or activity level. A quality specification based on biological variation or on state of the art was selected, considering the analytical performance of the available technology. (3) Maximum allowable differences (MAD) between analyzers were calculated. (4) Measurements were performed as stated in appendix B of the guideline. (5) Maximum differences between analyzers were calculated. Results were considered comparable when the maximum difference was less than or equal to the MAD.
Results
For the 24 parameters evaluated, any difference between analyzers exceeded the MAD.
Conclusions
The EP31-A-IR guideline proved to be useful for periodic verification of results’ comparability. However, it must be considered that, to be practicable, it may require to adjust the acceptance criteria in accordance to the analytical performance of the available technology; as well as the number of analytical measurements conforming to the laboratory resources.
-
Research ethics: Research involving human subjects complied with all relevant national regulations, institutional policies and is in accordance with the tenets of the Helsinki Declaration (as revised in 2013), and has been approved by the Ethics Committee of our Hospital (IIBSP-CED-2023-79).
-
Informed consent: Not applicable.
-
Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: The authors state no conflict of interest.
-
Research funding: None declared.
-
Data availability: The datasets generated during the current study are available from the corresponding author on reasonable request.
References
1. Clinical and Laboratory Standards Institute (CLSI). Measurement procedure comparison and bias estimation using patient samples. Approved guideline – CLSI document EP09-A3. Wayne, PA: CLSI; 2013.Search in Google Scholar
2. Braga, F, Panteghini, M. Commutability of reference and control materials: an essential factor for assuring the quality of measurements in Laboratory Medicine. Clin Chem Lab Med 2019;57:967–73. https://doi.org/10.1515/cclm-2019-0154.Search in Google Scholar PubMed
3. EN ISO 15189:2022. Medical laboratories – particular requirements for quality and competence. Geneva: ISO; 2022.Search in Google Scholar
4. Clinical and Laboratory Standards Institute (CLSI). Verification of comparability of patient results within one health care system. CLSI document EP31-A-IR. Wayne, PA: CLSI; 2012.Search in Google Scholar
5. Harris, EK. Statistical principles underlying analytic goal-setting in clinical chemistry. Am J Clin Pathol 1979;72:374–82.Search in Google Scholar
6. Fraser, CG, Petersen, PH. The importance of imprecision. Ann Clin Biochem 1991;28:207–11. https://doi.org/10.1177/000456329102800301.Search in Google Scholar PubMed
7. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM biological variation database. Available from: https://biologicalvariation.eu/ [Accessed 18 Aug 2023].Search in Google Scholar
8. Minchinela, J, Ricós, C, Perich, C, Fernández-Calle, P, Alvarez, V, Domenech, M, et al.. Biological variation database and quality specifications for imprecision, bias and total error (desirable and minimum). The 2014 update. Available from: https://www.westgard.com/biodatabase-2014-update.htm [Accessed 18 Aug 2023].Search in Google Scholar
9. de Medio, EP, Borrás, AM, Quiles, JA, Castellví, JB, Bauzá, FR, García, AS, et al.. Minimum analytical quality consensus specifications in Spain. Comparison with previous, mandatory and biological. Rev Med Lab 2020;1:93–107.Search in Google Scholar
10. Molina, A, Guiñon, L, Perez, A, Segurana, A, Bedini, JL, Reverter, JC, et al.. State of the art vs biological variability: comparison on hematology parameters using Spanish EQAS data. Int J Lab Hematol 2018;40:284–91. https://doi.org/10.1111/ijlh.12783.Search in Google Scholar PubMed
11. Lee, EJ, Lee, E, Kim, M, Kim, HS, Lee, YK, Kang, HJ. Verification of the comparability of laboratory results from two instruments within one health care system according to Clinical and Laboratory Standard Institute EP31-A-IR. J Lab Med Qual Assur 2016;38:129–36. https://doi.org/10.15263/jlmqa.2016.38.3.129.Search in Google Scholar
Supplementary Material
This article contains supplementary material (https://doi.org/10.1515/cclm-2023-0994).
© 2023 Walter de Gruyter GmbH, Berlin/Boston